Glyphosate is a non-selective herbicide which is adsorbed through leaves and was first sold by Monsanto under the Roundup trade name. This is one of the most widely used herbicides, regularly used for agriculture, horticulture, and silviculture applications.
Glyphosate is a non-selective herbicide which is adsorbed through leaves and was first sold by Monsanto under the Roundup trade name. This is one of the most widely used herbicides, regularly used for agriculture, horticulture, and silviculture applications. The United States Environmental Protection Agency (US EPA) requires that drinking water and raw surface water be monitored for the presence of glyphosate and related compounds using EPA Method 547.0. The European Union (EU) regulation (EC Directive 2005/70/EU) provides guidance with regards to the presence of glyphosate in drinking water supplies.
Instrument: |
Waters Alliance system for carbamate analysis |
Eluent: |
0.05% phosphoric acid |
Column: |
Ion Exclusion, 7.8 mm x 150 mm @ 55 °C |
Guard: |
Guard-Pak module and inserts |
Injection: |
200 μL of standard mix |
Flow Rate: |
1.5 mL / min |
Detection: |
Fluorescence, Ex- 340 nm, Em- 455 nm, Gain- 10 |
Refer to EPA Method 547 for preparation of field grab samples.
Filter through 0.45 μm Acrodisc. filters is described.
Oasis. MAX SPE Method for Glyphosate and Metabolite Method for Oasis MAX Cartridge, 6 cc, 150 mg
* Alternate eluent is 4 mL 0.6 M sodium citrate
Use 6 cc, 500 mg Oasis MAX for samples > 50 mL.
Pipette 100 μL of AccuStandard. mix (M-547) into 100 mL of acidified water for a concentration of 100. Prepare acidified water by adjusting the pH of HPLC grade water to 3.0 by dropwise addition of hydrochloric acid (HCL). Use EPA method 547-02 as above for AMPA (aminomethyl phosphonic acid).
Dilute 0.5 mL of 85% phosphoric acid (H3PO4) to 1 L, mix well, filter and degas.
Dissolve 1.35 g KH2PO4, 11.6 g NaCl, 0.4 g NaOH, and 0.2 mL Clorox. Bleach (plain) in water and dilute to 1 L, filter and degas.
Dissolve 0.8 g of OPA (o-phthdialdehyde) in 10 mL of methanol, add this to an aqueous solution of 19.1 g of Borax Na2B4O7•10H2O. Make to a final volume of 1 L, filter and degas. To this, add 2 mL of 2-mercaptoethanol, swirl gently to mix. Protect from light.
Note: Post-column flow rate for both reagents is 0.5 mL/min, post column reaction temperature is 38 °C. Insert second reaction coil in line before the fluorescence detector.
720002729, 2008