Quantitation of N-Nitroso-Propranolol in Drug Substance using LC-MS/MS
Abstract
N-Nitrosamines (nitrosamines) can be potentially mutagenic and carcinogenic. Since 2018, nitrosamines have been discovered in several classes of pharmaceuticals. There have been concerted efforts to mitigate the presence of nitrosamines in all marketed medicines while still ensuring continued safe access to vital medications.
Nitrosamine drug substance related impurities (NDSRIs) have been detected in multiple drug products. NDRSI formation occurs due to presence of susceptible structural properties, such as a secondary or tertiary amine, that allows for the addition of a nitroso group under certain chemical conditions, including the presence of nitrites from excipients.
Recently, both the EMA and the FDA provided revisions on how acceptable intake values (AI) are determined. The revisions introduced an expanded structure activity (SAR) approach with different potency categories based on carcinogenicity data derived from known nitrosamines (Carcinogenic Potency Categorization Approach, CPCA). In 2022, propranolol, a beta-blocker, was the subject of a recall due to the detection of the nitrosated impurity, N-nitroso-propranolol in drug product. This study describes the analysis of N-nitroso-propranolol in drug substance using Ultra-Performance Liquid Chromatography (UPLC) with electrospray detection (ESI) and a Xevo™ TQ-S micro Tandem Quadrupole Mass Spectrometer. Analytical methods used for the detection of NDSRIs, are required to have a lower limit of quantitation (LOQ) at 10% of the limit derived from the AI and the active pharmaceutical ingredients (API) maximum daily dose (MDD). Assuming a MDD of 320 mg, the assay exceeded the regulatory requirements according to the revised AI of 1500 ng/day for N-nitroso-propranolol. The method demonstrated good linearity over the concentration range of 0.01–100 ppm for N-nitroso-propranolol with less than 15% concentration deviations. The R2 was greater than 0.998. The method recovery ranged from 89.3–104.6% and the accuracy was within +/- 10% of the true value.
Benefits
- Trace level detection beyond the required regulatory limits as defined by new CPCA approach for the analysis of N-nitroso-propranolol using the Xevo TQ-S micro Tandem Quadrupole Mass Spectrometer in MRM acquisition mode
- Accurate and reproducible methodology for quantitative analysis of N-nitroso-propranolol in drug substance within the required recovery range
- Chromatographic resolution between propranolol, N-nitroso-propranolol and N-formylpropranolol
Introduction
Nitrosamines can be potentially mutagenic and carcinogenic.1–2 Since 2018, Nitrosamines have been discovered in several classes of pharmaceuticals.3–5 There have been concerted efforts to mitigate the presence of nitrosamines in all marketed medicines while still ensuring continued safe access to vital medications.6
Nitrosamine drug substance related impurities (NDSRIs), a term used to indicate the structural relationship between the active pharmaceutical ingredient (API) and its nitrosated impurity, have been detected in multiple drug products. NDRSI formation occurs due to presence of susceptible structural properties, such as a secondary or tertiary amine, that allows for the addition of a nitroso group under certain conditions, including the presence of nitrites from excipients.2
Recently, both the EMA and the FDA provided revisions on how acceptable intake values (AI) are determined.7–8 The revisions introduced an expanded structure activity (SAR) approach with different potency categories based on carcinogenicity data derived from known nitrosamines (Carcinogenic Potency Categorization Approach, CPCA). The calculation of a potency score aids in the assignment of the revised AI limits.9 The current potency categories range from 18 ng/day (EMA) and 26.5 ng/day (FDA) to 1500 ng/day. Flexibility in instrument sensitivity is an important feature to enable accurate measurements to be made for nitroso impurities across all potency categories. Tandem Quadrupole Mass Spectrometer analysers offer the best performance characteristics for quantitative measurements in complex matrices such as drug products due to their ease of use, robustness, sensitivity, and the assay selectivity afforded through selection of precursor and product ions in multiple reaction monitoring (MRM) experiments.
In 2022, propranolol, a widely used beta-adrenergic receptor blocker (beta-blocker) used in the treatment of hypertension, abnormal heart rhythms, and other cardiovascular disorders, was the subject of a recall due to the detection of the nitrosated impurity, N-nitroso-propranolol in drug substance using Ultra-Performance Liquid Chromatography (UPLC™) with electrospray detection (ESI) and a Tandem Quadrupole Mass Spectrometer. During the study, one of the API samples tested was found to contain N-formylpropranolol which has m/z 288 and fragments that in the absence of chromatographic separation can lead to signal overlap in some of the MRM transitions of N-nitroso-propranolol.
Experimental
LC Conditions
LC system: |
ACQUITY Premier Binary with an FTN |
Detection: |
PDA 210–400 nm: 230 nm single wavelength |
Vials |
TruView LC/MS (p/n: 186005663CV) |
Column: |
Xbridge® BEH C8 2.5 µm 3.0 x 100 mm (p/n: 186006047) |
Column temperature: |
45 °C |
Sample temperature: |
5 °C |
Injection volume: |
10 µL |
Flow rate: |
0.50 mL/min |
Mobile phase A: |
0.1% formic acid in water |
Mobile phase B: |
0.1% formic acid in methanol |
Extension loop: |
Extension Loop, 50-µL, HPS (p/n: 700012825) |
MS Conditions
Ionisation mode: |
Electrospray positive |
Capillary voltage (kV): |
1.0 |
Cone voltage (V): |
Table2 |
Collision energy (eV): |
Table 2 |
Source temperature (°C): |
130 |
Desolvation temperature (°C): |
500 |
Desolvation gas flow (L/hr): |
1000 |
Cone gas flow (L/hr): |
100 |
Sample Preparation
Standards and Reagents
Authentic standards of N-nitroso-propranolol were purchased from Toronto Research Chemicals (Toronto, Ontario). Propranolol hydrochloride was purchased from Sigma Aldrich (St. Louis, MO). N-Formylpropranolol was purchased from BOC Sciences (Shirley, NY). Optima LC/MS grade solvents and formic acid were purchased from Fisher Scientific. Primary stock solutions of propranolol (1 mg/mL), N-nitroso-propranolol and N-formylpropranolol at 10 mg/mL were used to prepare the working solutions.
Recovery, Accuracy and Precision Studies
Propranolol (C16H21NO2) constitutes 87.8% of propranolol hydrochloride (C16H21NO2 · HCl). Aliquots of propranolol hydrochloride (5.7 mg = 5.005 mg of propranolol) were weighted into 5 mL centrifuge tubes (Eppendorf). Spiking solutions were prepared in methanol at 2.5 µg/mL, 250 ng/mL and 25 ng/mL in glass scintillation vials. The spiking solutions were used to prepare three concentrations at which the recovery was evaluated.
Pre-spikes: 10 µL of each spiking solution concentration was pipetted into 6 tubes each containing 5.7 mg of propranolol hydrochloride. The samples were allowed to equilibrate for 30 min before 5 mL of methanol was added to make final concentrations of 5 ng/mL, 0.5 ng/mL and 0.05 ng/mL (n=6 at each concentration level). The centrifuge tubes were vortex mixed (1 min).
Post-spikes: Post spiked samples were prepared by weighing 5.7 mg of propranolol hydrochloride into 5 mL centrifuge tubes, 5 mL of methanol was then added. The three spiking solutions were used to prepare 6 replicates at each concentration level by adding 10 µL to the methanol to make 5 ng/mL, 0.5 ng/mL and 0.05 ng/mL concentrations. The centrifuge tubes were vortex mixed (1 min).
Matrix blanks: Matrix blanks were prepared by adding 5 mL of methanol to 6 tubes containing 5.7 mg of propranolol hydrochloride. The centrifuge tubes were vortex mixed (1 minute).
Standard solutions: Standard solutions at each test concentration level (5 ng/mL, 0.5 ng/mL and 0.05 ng/mL) (n=6) were prepared in 5 mL centrifuge tubes using 10 µL of the spiking solutions in 5 mL of methanol. The centrifuge tubes were vortex mixed (1 minute).
Calibration Curve Preparation
Authentic Standard Calibration Curve
An authentic standard of N-nitroso-propranolol (10 mg/mL) in methanol was sequentially diluted to create a calibration curve ranging from 0.005 ng/mL to 100 ng/mL.
Matrix Calibration Curve
Propranolol hydrochloride (21.34 mg) was weighted into a 20 mL glass scintillation vial. A stock solution of 1 mg/mL in propranolol was made up by dissolving the propranolol hydrochloride in 18.7 mL of methanol. This solution was used as the diluent for the N-nitroso-propranolol matrix calibration curve. An authentic standard of N-nitroso-propranolol was sequentially diluted using the 1 mg/mL propranolol solution to create a calibration curve ranging from 0.005 ppm to 100 ppm.
Results and Discussion
Chromatographic Separation
Photodiode array detector (PDA) data was collected simultaneously with the MRM data. The chromatographic resolution of the API, as well as other impurities detected in the UV, from the trace level impurities detected in the MRM data were evaluated based on the retention times (tR) observed. To avoid saturation of the MS the first 6.5 minutes of the chromatographic run was diverted to the waste using the integrated solvent divert valve. The separation between the propranolol API and the N-nitroso-propranolol was optimized on a C8 stationary phase (Table 1 and Figure 2). Chromatographic resolution of the API from the trace level impurities is important to avoid any potential matrix effects that could occur due to the proximity of the high levels of API present in the samples.
The MRM transitions were tuned using the Intellistart autotune feature present in the MassLynx™ software. Multiple MRM transitions can be useful for additional specificity if there is increased noise due to the matrix. Two of the optimized MRM transitions are shown in Figure 2, one of the confirmation transitions (289>259) and the quantifier transition (289>72).
Separation of N-formylpropranolol and N-nitroso-propranolol
Several API samples were tested during the study. In one sample, the N-formylpropranolol impurity with its structure shown in Figure 1, was detected. The identity of N-formylpropranolol was verified using an authentic standard. N-Formylpropranolol has an [M+H]+ of m/z 288, N-nitroso-propranolol has an [M+H]+ 1 amu higher at m/z 289. Two of the optimized MRM transitions for N-formylpropranolol, 288>144 and 288>102 are one amu less in both the precursor and the product ion masses when compared to the confirmatory transitions of N-nitroso-propranolol (Table 2). It follows that the 13C isotope of N-formylpropranolol will give rise to a signal in two of the MRM transitions used for N-nitroso-propranolol 289>145 and 289>103 since the 13C will be one amu higher. Figure 3 shows a chromatographic peak at the same tR for N-formylpropranolol (7.14 min) in the 289>145 MRM transition for N-nitroso-propranolol. Consequently, to avoid signal overlap of these MRM transitions, chromatographic resolution is recommended. Using the proposed method, it is possible to resolve N-formylpropranolol from N-nitroso-propranolol (Figure 3).
Chromatographic Peak Shape
An injection solvent containing high levels of organic can lead to deleterious solvent effects resulting in degradation of the chromatographic peak shape. The injection solution of 100% methanol used for this assay leads to peak shape distortion. However, the peak shape can be dramatically improved by adding some pre-column volume to the flow path. An extension loop (50 µL volume, p/n: 700012825) was added to port 6 of the Sample Manager valve allowing the preservation of the impurity peak shape (Figure 4).
Limit of Detection (LOD), Limit of Quantitation (LOQ) and Linear Dynamic Range
The quantitative limits of the assay were initially established using authentic standards of N-nitroso-propranolol. The LOD and LOQ based on 3:1 and 10:1 signal-to-noise (S/N) respectively, are shown in figure 5. The LOD and LOQ for an authentic standard of N-nitroso-propranolol were found to be 0.005 ng/mL and 0.01 ng/mL respectively.
Establishing the LOD and LOQ of N-nitroso-propranolol in the propranolol API was complicated by the presence of endogenous levels of the impurity in each of the propranolol API samples tested. The lower levels of the calibration curve made up in 1 mg/mL propranolol are shown in Figure 6.
Table 3 summarizes the average peak area responses obtained for 6 replicate injections of the API samples (1 mg/mL propranolol) spiked at 0.005 ppm and 0.01 ppm with N-nitroso-propranolol. The average peak area responses for authentic standards of N-nitroso-propranolol at the LOD of 0.005 ng/mL and LOQ 0.01 ng/mL are also shown. The responses for the N-nitroso-propranolol in the spiked API samples were corrected for the endogenous level of impurity using the average peak response of the N-nitroso-propranolol present in the API (n=6). The corrected peak area response of the spiked samples was then compared with the response of the authentic standards. The corrected responses at the LOD and LOQ in the API are within 10% of the responses observed in an authentic standard at the same concentration levels.
The threshold level of N-nitroso-propranolol is 4.69 ppm, 10% of the threshold level is 0.469 ppm. The response obtained for N-nitroso-propranolol spiked at 5 ppm and 0.5 ppm in 1 mg/mL of propranolol, and the API matrix blank are shown in Figure 7 along with the S/N calculated using the PtP algorithm, illustrating the ability of the method to exceed the sensitivity required at the regulatory threshold levels.
An authentic standard of N-nitroso-propranolol was used to determine the linear dynamic range before evaluating the N-nitroso-propranolol in the presence of the API. The curve, when injected in triplicate was linear from 0.005-100 ng/mL with and R2 of 0.998 and <10% concentration deviations across the range (data not shown).
The calibration curve for the N-nitroso-propranolol in 1 mg/mL of the API is shown in Figure 8. The R2 was 0.998 and the concentration deviations were less than 15% over the range of 0.01 to 100 ng/mL (4 orders of magnitude) using 1/X weighting.
Recovery, Precision and Accuracy
The recovery experiments were performed at 0.05 ppm, 0.5 ppm and 5 ppm (n=6). The results are summarized in Table 4. Due to the presence of the endogenous level of N-nitroso-propranolol in the API, the recovery calculations were based on the corrected responses for the pre-spiked API samples using the average response for the N-nitroso-propranolol peak present in the API blank (n=6). The average response for the peak in the API was subtracted from each of the responses in the pre-spiked samples. The recovery was calculated using the following formula:
% Recovery = (Response Pre-spike corrected/Standard Response) *100%
The calculated recoveries were between 89.3–104.6%. The %RSD for the recoveries at each concentration level were less than 5%.
Method precision and accuracy was evaluated at 0.5 ppm and 5 ppm (Table 5). The precision of the assay was measured by the %RSD of 6 replicate measurements made at 0.5 ppm and 5 ppm. The %RSD were less than 2% for the levels tested. The accuracy was measured as the percentage deviation from the nominal value and was accurate to within 10%.
Matrix effects were assessed at 0.05 ppm, 0.5 ppm and 5 ppm by comparing the response obtained in neat standards to the corrected response for the post spiked samples. The matrix effects were found to be <10% for all levels tested.
Conclusion
In the presented study, a UPLC-MS/MS method was developed for the quantitation of N-nitroso-propranolol in drug substance. The method can perform high sensitivity quantitative analysis, with good precision, and accuracy within the required recovery range to meet regulatory guidelines. In addition, the impurity, N-formylpropranolol can be chromatographically resolved from N-nitroso-propranolol avoiding the overlap of the signals from the 13C isotopes in the N-nitroso-propranolol MRM transitions (289>145 and 289>103). The chromatographic resolution is important to retain specificity of the assay thus reducing the risk of misreporting impurity levels.
References
- Montesano, R.; Hall, J. Nitrosamine Metabolism and Carcinogenesis. In Mutation, Cancer, and Malformation, Chu, E. H. Y., Generoso, W. M. Eds.; Springer US, 1984; pp 447–464.
- Holzgrabe, U. Nitrosated Active Pharmaceutical Ingredients - Lessons Learned? J Pharm Sci 2023, 112 (5), 1210–1215. DOI: 10.1016/j.xphs.2023.01.021 From NLM.
- Moser, J.; Schlingemann, J.; Saal, C. N-Nitrosamines Impurities in Pharmaceuticals The Abrupt Challenges that Resulted, the Evolving Science, and the Regulatory Framework. Journal of Pharmaceutical Sciences 2023, 112 (5), 1161–1162. DOI: https://doi.org/10.1016/j.xphs.2023.01.016.
- Teasdale, A.; Moser, J.; Ford, J. G.; Creasy, J. N -Nitrosamines. In Mutagenic Impurities, Teasdale, A. Ed.; Wiley, 2021; pp 269–320.
- Bharate, S. S. Critical Analysis of Drug Product Recalls due to Nitrosamine Impurities. Journal of Medicinal Chemistry 2021, 64 (6), 2923–2936. DOI: 10.1021/acs.jmedchem.0c02120.
- Charoo, N. A.; Dharani, S.; Khan, M. A.; Rahman, Z. Nitroso Impurities in Drug Products: An Overview of Risk Assessment, Regulatory Milieu, and Control Strategy. AAPS PharmSciTech 2023, 24 (2), 60. DOI: 10.1208/s12249–023–02523–w.
- Nitrosamines EMEA-H-A5(3)-1490 - QA Art. 5(3) Implementation_ QA10_revision 18_Oct23 (europa.eu) [accessed October 16th, 2023].
- Updated Information | Recommended Acceptable Intake Limits for Nitrosamine Drug Substance-Related Impurities (NDSRIs) | FDA [accessed October 16th, 2023].
- Burns, M. J.; Ponting, D. J.; Foster, R. S.; Thornton, B. P.; Romero, N. E.; Smith, G. F.; Ashworth, I. W.; Teasdale, A.; Simon, S.; Schlingemann, J. Revisiting the Landscape of Potential Small and Drug Substance Related Nitrosamines in Pharmaceuticals. Journal of Pharmaceutical Sciences. DOI: 10.1016/j.xphs.2023.10.001 (accessed October 16th, 2023).
- Pfizer Recalls Inderal-LA (propranolol hydrochloride) Capsules due to a Nitrosamine Impurity. Pfizer Recalls Inderal-LA (propranolol hydrochloride) Capsules due to a Nitrosamine Impurity - Canada.ca. [accessed October 16th, 2023].
- Li, X.; Le, Y.; Seo, J.-E.; Guo, X.; Li, Y.; Chen, S.; Mittelstaedt, R. A.; Moore, N.; Guerrero, S.; Sims, A.; et al. Revisiting the Mutagenicity and Genotoxicity of N-nitroso Propranolol in Bacterial and Human In Vitro Assays. Regulatory Toxicology and Pharmacology 2023, 141, 105410. DOI: https://doi.org/10.1016/j.yrtph.2023.105410.
720008171, January 2024