For forensic toxicology use only.
This application note details the extraction and UPLC-MS/MS analysis of THC and its metabolites from urine using Oasis PRiME μElution Plates.
The use of Oasis PRiME resulted in consistent and highly reproducible recoveries of all compounds with minimal matrix effects. The μElution format allowed for the concentration of the sample on the SPE column, eliminating the need to evaporate and reconstitute the sample, minimizing the risk of analyte loss due to nonspecific binding and streamlining the laboratory workflow. This resulted in a method that was linear, accurate and precise for all analytes, with limits of quantification of 0.1 ng/mL for THC and its metabolites.
Sample preparation is an important consideration for any bioanalytical LC-MS/MS method designed for forensic toxicology. Waters has developed a novel sample preparation sorbent, Oasis PRiME, which is designed to have some key advantages over traditional SPE sorbents. These include the ability to eliminate sorbent preconditioning and equilibration, allowing a more rapid workflow compared to traditional SPE products, and the ability to remove more interferences, resulting in a cleaner extracts and reducing the risk of short column lifetimes or MS source fouling.
This application note details the extraction and UPLC-MS/MS analysis of Δ-9-tetrahydrocannabinol (THC) and its metabolites, 11-hydroxy- Δ-9-THC (THC-OH) and 11-nor-9-Carboxy-Δ-9-THC (THC-COOH) from urine using Oasis PRiME μElution Plates. Δ-9-tetrahydrocannabinol (THC) is the main psychoactive element present in the plant Cannabis sativa. Quantitative analysis of these compounds in urine is an indicator of cannabis consumption, with high levels indicating recent and/or chronic use.
The use of Oasis PRiME resulted in consistent and highly reproducible recoveries of all compounds with minimal matrix effects. The μElution format allowed for the concentration of the sample on the SPE column, eliminating the need to evaporate and reconstitute the sample, minimizing the risk of analyte loss due to nonspecific binding and streamlining the laboratory workflow. This resulted in a method that was linear, accurate and precise for all analytes, with limits of quantification of 0.1 ng/mL for THC and its metabolites.
All standards and stable isotope labelled internal standards were purchased from Cerilliant (Round Rock, TX, USA). Stock standards at 100 μg/mL were prepared in 40% methanol (THC, THC-OH, and THC-COOH). A working internal standard solution of 1 μg/mL THC-D3, THC-OH-D3 and THC-COOH-D3 was also prepared in 40% methanol. Individual calibrators and quality control standards were prepared daily in 40% methanol. 80 μL of each working calibrator or QC standard was added to 1920 μL of human urine to make calibration curves and QC samples.
β-Glucoronidase from E. Coli K 12 was purchased from Roche Life Science (Indianapolis, IN)
Glucuronide hydrolysis: 40 μL internal standards was added to 2 mL spiked human urine sample in a glass vial, then 2.4 mL 0.1 M potassium phosphate buffer (pH 6.8) containing 10 μL β-Glucoronidase was added. Vials were capped, vortex mixed, and incubated at 37 °C water bath for 16 hours. After allowing samples to cool down to room temperature, 150 μL of 10 M NaOH was added, vortex mixed and hydrolyzed in a dry heating block for 30 min at 70 °C. Once the samples had cooled, 850 μL glacial acetic acid was added to the samples and vortex mix
Solid-Phase Extraction with Oasis PRiME μElution Plate: 500 μL pretreated sample (equivalent to 180 μL urine) was directly applied to the Oasis PRiME μElution Plate. All wells of the SPE plate were then washed with 2 x 300 μL aliquots of 25% methanol. The samples were then eluted with 2 x 25 μL aliquots of 60:40 ACN:IPA and diluted with 50 μL of water. 5 μL was injected onto the UPLC-MS/MS system. The SPE extraction procedure is summarized in Figure 1.
Analyte recovery was calculated according to the following equation:
Where A equals the peak area of an extracted sample and B equals the peak area of an extracted blank matrix sample in which the compounds were added post-extraction.
Matrix effects were calculated according to the following equation:
The peak area in the presence of matrix refers to the peak area of an extracted matrix sample in which the compounds were added post-extraction. The peak area in the absence of matrix refers to analytes in a neat solvent solution.
LC system: |
ACQUITY I-Class UPLC System |
Column: |
ACQUITY UPLC BEH C18 Column, 130Å, 1.7 μm, 2.1 x 50 mm |
Column temp.: |
40 °C |
Sample temp.: |
10 °C |
Mobile phase A (MPA): |
Water with 0.1% formic acid |
Mobile phase B (MPB): |
ACN with 0.1% formic acid |
Strong wash solvent: |
70:30 ACN:Water with 2% formic acid |
Time(min) |
Flow(mL/min) |
%A |
%B |
---|---|---|---|
0 |
0.6 |
50 |
50 |
1.0 |
0.6 |
50 |
50 |
3.0 |
0.6 |
5 |
95 |
3.5 |
0.6 |
5 |
95 |
3.6 |
0.6 |
50 |
50 |
4.0 |
0.6 |
50 |
50 |
MS system: |
Xevo TQ-S Mass Spectrometer |
Ionization mode: |
ESI Positive |
Capillary voltage: |
2.0 kV |
Cone voltage: |
Optimized for each analyte |
Desolvation gas: |
1000 L/hr |
Cone gas: |
150 L/hr |
Desolvation temp.: |
500 °C |
Source temp.: |
150 °C |
Data were acquired and analyzed using MassLynx Software (v4.1). Quantification was performed using TargetLynx.
Figure 2 shows chromatography of the three cannabinoids from an extracted calibrator at 2 ng/mL. All compounds eluted within 3 minutes with all peak widths were under 3 seconds at 5% of baseline. All peaks were symmetrical with symmetries between 0.95–1.15.
Table 2 lists the retention time and individualized MS parameters of the cannabinoids and their stable isotope labelled internal standards, including MRM transitions, cone voltage, and collision energy. Two MRM transitions were used for each compound, a primary (listed first) and a confirmatory transition (listed second).
Extraction recoveries were very consistent. As Figure 3 shows, recovery for THC-OH and THC-COOH was around 90% and THC was 60% with all %RSDs under 7.5%, demonstrating the reproducibility of Oasis PRiME. Matrix effects were minimal, at less than 15% for all compounds. Once again, the low standard deviations (7.5% or less) and high recoveries for THC-OH and THC-COOH demonstrate the consistency of extraction and cleanup seen with Oasis PRiME HLB. All recovery and matrix effect data are summarized in Table 3. Oasis PRiME HLB also provided better recovery, variability and matrix effects than LLE, with a more simplified procedure.1
Calibration and quality control samples were prepared as previously described in the materials and method section. Calibration ranges were from 0.1–100.0 ng/mL for THC-COOH and THC-OH and 0.2–100.0 ng/mL for THC. Quality control samples were prepared at low, medium, and high concentrations as appropriate for the calibration ranges.
All compounds had linear responses over the entire calibration range with R2 values of 0.99 or greater with 1/x weighting. Table 4 summarizes the data from the calibration curves. Lower limits of quantification (LLOQ) were 0.1 ng/mL for THC-COOH and THC-OH and 0.2 ng/mL for THC. In each case, all FDA recommendations for accuracy, precision, linearity and analytical sensitivity were met for validated methods.2
Quality control samples were accurate and precise. All results were within 15% of expected values and %RSDs were under 2% (N=6). This data can be seen in Table 5. The excellent accuracy and precision demonstrate the consistency and robustness of this sorbent.
Calibration and quality control samples were prepared as previously described in the materials and method section. Calibration ranges were from 0.1–100.0 ng/mL for THC-COOH and THC-OH and 0.2–100.0 ng/mL for THC. Quality control samples were prepared at low, medium, and high concentrations as appropriate for the calibration ranges.
All compounds had linear responses over the entire calibration range with R2 values of 0.99 or greater with 1/x weighting. Table 4 summarizes the data from the calibration curves. Lower limits of quantification (LLOQ) were 0.1 ng/mL for THC-COOH and THC-OH and 0.2 ng/mL for THC. In each case, all FDA recommendations for accuracy, precision, linearity and analytical sensitivity were met for validated methods.2
Quality control samples were accurate and precise. All results were within 15% of expected values and %RSDs were under 2% (N=6). This data can be seen in Table 5. The excellent accuracy and precision demonstrate the consistency and robustness of this sorbent.
720005556, December 2015