SFC Prep 150 AP System

SFC Prep 150 AP System

Designed for automated UV and/or MS-triggered purification

Designed for automated UV and/or MS-triggered purification

The SFC Prep 150 AP System is the first and only commercially available preparative SFC system offering automated sample handling, column switching, fraction collection and tracking in an easy-to-use, open bed format.

The SFC Prep 150 AP System provides routine compound purification for drug discovery labs using the MassLynx Software with FractionLynx Application Manager, offering fast separations, high resolution, and high throughput. The inherent benefits of SFC, combined with Waters’ world-renowned service and support provides users with a robust, cost-effective solution to any purification lab.

SFC Prep 150 AP System. SFC Prep 150 AP System

Specifications

Maximum pressure range

300 bar (4350 psi)

Cooling

Circulating coolant

Total flow rate range (CO2 + Co-solvent)

20.0 to 150 mL/min

Recovery

No Ketoprofen detected, therefore carry over is less than 0.01%

Individual component specifications may allow operation outside the total system defined parameters. The system is designed and tested only for the ranges defined in system specifications.

Ambient operating temperature

15 to 40 °C

Ambient operating humidity

20 to 80% relative humidity at 40 °C

Pressure rating

10,000 psi (680 bar)

Maximum operating pressure

8700 psi (592 bar)

Flow range

20 to 200 g/min

Wetted materials

316 stainless-steel, 440C, 17-4PH stainless-steel, sapphire, GFPM, PTFE, AL

Number of solvents

Four solvents selected by the Solvent Select Valve (SSV)

Solvent conditioning

Sparging one channel per solvent

Operating flow rate range

0.50 to 150.00 mL/min, in 0.01 mL/min increments

Maximum operating pressure

6000 psi up to 100 mL/min, roll-off to 5000 psi at 150 mL/min

Plunger seal wash

Integral, active, programmable

Delay volume

<6.5 mL

Compressibility compensation

Automatic and continuous

Priming

Self prime from solvent container on floor with dry lines

Compositional accuracy

±3% absolute (full scale) from 5% to 95%; and from 1 to 150 mL/min (methanol or water step gradient) at 600 psi

Compositional precision

Retention time variation <0.15 min SD (uracil/caffeine; 1, 5, 20, 50, 150 mL/min; 70:30 water/methanol dial-a mix)

Sample capacity

Injection: Up to 15 microplates (up to 384 well)

Up to 72 vials

Up to 544 test tubes

Collection: Up to 480 test tubes

Up to 12 deep-well plates

Up to 128 28 mm vials

Syringe delivery accuracy

>99.0% at full stroke with 1000 µL syringe

Carryover

<0.05%

Injection volume

Preparative: 1 mL standard (fixed loop), (0.024 in valve passages)

System pressure ratings

6000 psi High pressure side, including injector

100 psi Collection side and syringe/needle wash

Wash pump

>30 mL/min flow through standard tubing and needle; two wash solvents.

Cycle controlled by “Wash Factor” setting

Available loops

1.0 mL (standard preparative) 2.5 mL, 5.0 mL, 12.5 mL, 30.0 mL

Supported syringes

1.0 mL (standard) 2.5 mL, 5.0 mL, 12.5 mL

Injection cycle time

<30s based on standard wash cycle

Control software

MassLynx v4.2 or higher

Physical/environmental

Without hood

Width: 959 mm (37.75 in)

Depth: 629 mm (24.75 in)

Height: 1020 mm (40 in)

With hood

Width: 959 mm (37.75 in)

Depth: 629 mm (24.75 in)

Height: 1060 mm (41.75 in)

Default syringe

5 mL

Default loop size

2 mL, (Optional 5 mL, 10 mL)

Injection volume range

0.010 mL (10 µL) to 9.5 mL (9500 µL)

Sample delivery precision

Less than 1.0% RSD; 0.75 mL injection volume, 2.0 mL loop

Injection linearity

R2 greater than 0.995 using 20–75% of loop volume (2 mL loop)

Sample carryover

No Ketoprofen detected, therefore carry over is less than 0.01%

Primary wetted materials

316 stainless-steel, PEEK, PTFE, borosilicate glass

The Prep Column Oven is a thermally controlled module designed for use with preparative SFC columns.

Temperature range

Ambient +5.0 to 70.0 °C

Maximum heat rate

Approx 6.0 °C/min

Temperature accuracy

±0.5 °C

Maximum column specification

1 to 6, 20 x 250 mm columns

Wavelength range

190 to 800 nm

Light source

Prealigned, Intelligent Technology Deuterium lamp

Wavelength accuracy

±1.0 nm

Linearity range

<5% at 2 AU (Propylparaben 257 nm, 10 mm cell)

Optical resolution

1.2 nm

Noise, wet

≤60 μAU (254 nm, 2 Hz, 1 s TC, 3.6 BW res, 10 mm analytical cell)

Drift, dry

≤5000 μAU/h (2 h warm-up, constant temperature and humidity at 230 nm, 3.6 BW res, 2 Hz)

Data rate

Up to 80 Hz

Path length

3 mm (semi prep)

Cell volume

18.3 μL (semi prep cell)

Pressure limit

6000 psi (408 bar)

Wetted materials

316 stainless-steel, PEEK, fused silica

Line voltage

100 to 240 VAC

Number of solvents

One

Solvent conditioning

One integrated, vacuum degassing channel

Primary and accumulator check valves

Passive

Pump seal wash

Equipped with a wash system to flush the rear of the high pressure seal and the plunger

Compressibility compensation

Automated and continuous with no user intervention required

Priming

Automated priming at flow rates of 4 mL/min

Flow accuracy

1.0% of set flow at 0.5 to 2.0 mL/min. Backpressure: 1000 psi ±20% with degassed MeOH using either a volumetric or mass based method

Flow precision

<0.075 %RSD or <0.02 minutes SD, whichever is greater based on six replicate injections

Acquisition modes

Full scan MS

Selected Ion Recording (SIR)

RADAR

An information-rich acquisition approach that allows you to collect highly specific quantitative data for target compounds while providing the ability to visualize all other components

Mass range

30 to 1250 m/z

Scan speed

Automatically-optimized for enhanced data quality for acquisition rates of up to e.g. 10 Hz for m/z 100 to 1000 or 20 Hz for m/z 50 to 500

Mass accuracy

Mass accuracy is better than ±0.2 Da over full mass range

Mass stability

Mass drift is less than 0.1 Da over a 24 hour period

Linearity of response

The linearity of response relative to sample concentration, for a specified compound, is up to four orders of magnitude from the limit of detection

Ion polarity switching time

25 ms to switch between positive and negative ion modes

SIR acquisition rate

Automatically-optimized for enhanced data quality for acquisition rates of up to 100 Hz

Number of SIR channels

Up to 1024 SIR channels (32 functions, 32 channels per function) can be monitored in a single acquisition

Mass resolution

Automated mass resolution control (0.7 Da) for constant data quality

SIR sensitivity (ESI+)

A 100 pg on column injection (5 μL of 20 pg/μL) of sulfadimethoxine, with a mobile phase flow rate of 800 μL/min will give a chromatographic signal-to-noise for m/z 311 greater than 2000:1 (400:1 with integrated diaphragm backing pump)

SIR sensitivity (ESI-)

A 50 pg on column injection (5 μL of 10 pg/μL) of chloramphenicol, with a mobile phase flow rate of 800 μL/min will give a chromatographic signal-to-noise for m/z 321 greater than 300:1 (60:1 with integrated diaphragm backing pump)

SIR signal-to-noise definition

Signal is defined as the height of the chromatographic peak of interest and noise is defined as the RMS of a continuous section of the mass chromatogram

The ABPR valve assembly is motor driven and temperature controlled for cooling during depressurization. A built-in pressure sensor provides closed loop feedback for control and pressure alarm monitoring.

Maximum flow rate

150 mL/min

Maximum pressure

4400 psi (300 bar)

After exiting the Automated Back Pressure Regulator, this heat exchanger is used to ensure a phase separation in the mobile phase between the CO2 and the co-solvent. This “trailing” Heat Exchanger is used to vaporize the CO2 from the mobile phase so that the Gas Liquid Separator is collecting only a sample-enriched solvent mixture.

Pressure rating

10,000 psi (680 bar)

Wetted material

316 stainless-steel

Collecting in the same tubes or racks is achieved by eliminating the majority of the carbon dioxide, directing only the liquid phase along with the collected fraction to the collector.

Total flow rate range

20 to 150 mL/min

Co-solvent max flow range

75 mL/min

Operational pressure range

(Optimal)

0 to 100 psi (0 to 6.8 bar)

50 psi (3.4 bar)

Wetted materials

316 stainless-steel, PEEK, borosilicate glass

External control

MassLynx v4.2

External communication

Ethernet and RS-232 communications

Event input/outputs

Rear panel contact closure and/or TTL inputs/outputs

Operating system

Windows 10

Overview

  • Offers automated sample handling, column switching, fraction collection, and tracking in an easy-to-use, open bed format
  • Achieve fast separations, high resolution, and high throughput
  • Attain routine compound purification for drug discovery labs, using the Waters MassLynx/FractionLynx Application Manager
  • Benefit from a robust, cost-effective solution to any purification lab, when combined with world-renowned service and support from Waters

Recommended Use: For drug discovery labs needing fast separations, high resolution, and high throughput in routine compound purification.


Features Header

Features Header






Automate the purification process

The SFC Prep 150 AP System automates the purification process by linking analytical pre-screen, purification and fraction re-analysis with AutoPurify in an “open access” setting.

SFC systems utilize liquid carbon dioxide (CO2) as its main mobile phase in combination with one or more organic solvents resulting in faster equilibration, lower pressure drops across the column, solvent reduction, and decreased cost per sample. The process is reproducible and applicable to a wide range of compounds relevant in the pharmaceutical, life sciences, chemical materials, food, and environmental markets.


Moving SFC forward

The SFC Prep 150 AP System comes equipped with modules to help with routine compound purification, including:

  • QGM Low Pressure mixing Quaternary Gradient Pump: This pump delivers flowrates up to a maximum of 150 mL/min as serves as the co-solvent delivery device for this system.
  • P200X CO2 Pump: This is a high-pressure delivery pump having dual stainless-steel heads with a cam driven sapphire piston assembly; self-priming check valves; presser sensor; pressure gauge; brushless motor; and a rupture disc assembly. The pumps design lends itself to control based upon feedback from the pressure sensor and mass flowmeter.
  • 3767 Sample Manager: This module provides injection and collection capability in an open bed format. The system includes the fume hood assembly for the 3767 for CO2 management. Injection and collection racks must be ordered for the system. The specific racks required will be determined by your needs and requirements.
  • Analytical-2-Prep Column Oven: A thermal control module, the Analytical-2-Prep Column Oven's unique drawer design gives the user unprecedented flexibility, enabling the use of 10 mm, and 20 mm diameter columns simultaneously.

Optional modifier stream injection

The Modifier Stream Injector’s function is to inject one sample at a time directly into the co-solvent flow. This device allows for settable air gaps on either side of the sample, which function as a buffer between the solvent and the sample decreasing dilution. There are two benefits to the MSI:

  • The sample is injected into the organic portion of the mobile phase (co-solvent) prior to mixing with CO2. The idea is to mitigate the diluent affect by introducing the sample without affecting the overall strength of the mobile phase, and maintaining the programmed solvent percentages throughout the injection and run. Modifier-stream injection provides improved peak shape and resolution, allowing for larger injection volumes, and higher loading (image 1).
  • It allows for Stacked Injections. Stacked injections reduce the time between injection cycles and minimize solvent usage. Stacked injections also significantly improve throughput by utilizing all of the available chromatographic space for continuous separation and purification. Usually, injections are made while an already injected sample is on (or eluting from) the column (image 2).


Resources

Documents

Documents


Support

Support



Related

The 3767 Sample Manager is a high-capacity sample processing system that easily and automatically manages concurrent sample injection, collection of peak fractions, and fraction analysis on a single HPLC or LC-MS platform.

ACQUITY QDa II Detector is a mass detector built around the needs of analytical scientists for chromatographic analysis. Robust, reliable, and requiring no sample adjustments, it integrates with your current LC, ACQUITY UPLC, UPC2, and purification systems.

Increase throughput and flexibility in preparative liquid chromatography with the extensive portfolio of purpose-built Waters purification solutions and systems.
Back To Top Back To Top