Mass Accuracy & Resolution

Mass Accuracy & Resolution

Increased, measured mass accuracy and resolution is now a dominant tool for structural characterization in various applications beyond early drug discovery. With their broad reach of specificity and utility, quadrupole time-of-flight (QTOF) instruments, offered by a number of manufacturers today,  are replacing other LCMS technologies.

Even though higher order instruments exist, a QTOF instrument’s high mass accuracy falls within a few parts per million of the true, calculated, monoisotopic value, and its high resolution—as much as 10 times higher than a quadrupole instrument’s—permits us to determine empirical formulas according to mass defect (where the critical mass value of hydrogen and other atoms present serve as a differentiator). Speciation analysis—discerning the difference between an aldehyde and a sulphide, for example—becomes possible with an increase in mass accuracy above the quadrupole limits to 30 ppm, where the two masses differ by 0.035 Da. Differentiation between the metabolic processes involving methylation is more demanding, however. Adding CH2 produces an increase over the precursor (response for the drug alone) in the measured mass of +14.0157 Da, as compared with a two-stage biotransformation involving hydroxylation (addition of oxygen) followed by oxidation at a double bond (loss of H2), which produces an increase of +13.9792 Da. Yet both measurements, when limited by nominal resolution—a typical quadrupole response—will look like +14 Da.

High mass-accuracy and low resolution

Low resolution quadrupole instruments perform well for extremely high mass-accuracy measurements, like those used for analyzing proteins. The masses of proteins are generally defined as “average” values when the isotope peaks are not resolved relative to each other. Average mass is the weighted mean of all the isotopic species in a molecule. The instrumental resolution normally employed on quadrupole instruments broadens the resolved response for a 10 kDa protein by a factor of x1.27. That factor increases significantly as the mass increases (for example, to x2.65 at 100 kDa). However, by reducing the peak width to m/z 0.25 (increasing resolution to 4000 resolution) rather than limiting the instrument resolution to 1000 using the typical peak width (m/z 0.6) improves the situation dramatically.

In practice, ESI-MS analyses of large molecules produce multiply charged ions. Hence the widths need to be divided by the number of charges on an ion to give the width on the mass-to-charge ratio scale. For example, a 20 kDa protein with 10 or 20 charges on it will produce isotope envelopes that are 0.9 or 0.45 m/z units wide at m/z ~2000 or ~1000 respectively.

When these ions are observed on an instrument set for a significantly lower resolution than that required to resolve the isotopes (say less than 10,000 resolution), a single peak is produced for each charge state. The overall width is determined by combining the instrument peak width with the theoretical width of the isotopic envelope divided by the number of charges on the ion. The instrumental peak width would be determined on the first isotope peak of a low-molecular-weight compound at the same m/z value as the multiply charged protein peak.

How much accuracy do we need, or can realistically achieve, and what are the compromises?

Consider the requirements for unambiguous characterization from the Journal of the American Society for Mass Spectrometry author’s guidelines (March 2004). For C, H, O, N compositions (C0-100, H3-74, O0-4 and N0-4) a nominal mass-to-charge response at 118 needs only an error not exceeding 34 ppm to be unambiguous, where a m/z response at 750 requires precision better than 0.018 ppm to eliminate “all extraneous possibilities.”

The effect of increasing mass accuracy for unambiguous identification of compounds (Quenzer, T.L., Robinson, J.M., Bolanios, B., Milgram, E. and Greig, M.J., Automated accurate mass analysis using FTICR mass spectrometry, Proceedings of the 50th Annual Conference on Mass Spectrometry and Allied Topics, Orlando, FL, 2002).
Increased filtering or restriction of error in the measurementreduces the possible candidates for a given result.

Comparing precision from instrument to instrument: millimass units (mmu), measurement error (ppm), and resolution

According to the Accurate Mass Best Practice Guide of the VIMMS Programme, an initiative that forms part of the UK National Measurement System, most instruments used for accurate mass measurements are capable of achieving precision of 10 ppm or better.

A calculated mass of 118 Da measured by a modern mass spectrometer to within 2 mmu accuracy would display 17 ppm error, sufficient by today’s standards for unambiguous determination of a chemical formula of that mass:

Monoisotopic calculated exact mass = 118 Da

Measured accurate mass = 118.002 Da

Difference = 0.002 mmu

Error [Difference/exact mass x 106] = 17 ppm

An instrument capable of a response at 750 m/z, also deficient by 2 mmu, would be accurate to 2.7 ppm. In the first case, the measurement is more than sufficient for unambiguous identification of a chemical formula, according to the published standards of the Journal of The American Society for Mass Spectrometry. But in the latter case, the measurement is insufficiently precise. Only the highest order Fourier transform ion cyclotron resonance mass spectrometry (FTICR) can achieve such precision at higher masses.

A comprehensive method of evaluating instrument mass accuracy measurement capability which resembles intended use is by calculating the root mean square or RMS error. To illustrate its use, the following is adapted from the mass measurement accuracy specification of a commercial TOF mass spectrometer.

"The mass measurement accuracy of the instrument, under normal operating conditions, will be better than a given ppm RMS over the given m/z range, based on a number of consecutive repeat measurements of an analyte peak (of given m/z), using a suitable reference peak (of given m/z). Analyte and reference peaks must have sufficient intensity and be free of interference from other masses."

There are some important points and assumptions need to be considered:

  1. An instrument calibration has already been performed with peaks of known mass using a calibration standard. The reference peak is used to account for any variation in the instrument calibration over time and mass measurement accuracy is determined using the analyte peak.
  2. Normal operating conditions - can also include details of chromatographic conditions (for LC-MS performance specifications) and any related MS operating conditions (e.g. mass resolution, m/z of interest, or spectral acquisition rate).
  3. Sufficient intensity - assumes that the ion count is not detrimental to characterization of the (mass measurement) accuracy and precision of the instrument in question. Too few ions leads to poor ion statistics and too many ions can lead to detector saturation, both of which result in a greater variation in the standard deviation of repeat measurements and will adversely influence calculation of the RMS error (also relevant to instrument calibration).
  4. Free from interferences - assumes that the mass measurement of the peak of known mass is free from interference by ions of the same or similar mass. Overlapping peaks lead to poor mass measurement accuracy which is also detrimental to properly characterizing the accuracy or precision of the instrument (also relevant to instrument calibration).
  5. The reference is a good representation of the m/z range which is relevant to the analysis of a particular sample type.

The RMS error is calculated using the following relation, where Eppm is the ppm error, and n is the number of masses considered:

It is worth noting the RMS error allows some measurements to fall outside the ppm error “window of interest” (e.g., 5 ppm RMS). To ensure quality measurements, the conditions described above need to be satisfied (particularly regarding intensity and influence of interferences - balanced ion statistics with clear peak definition in the spectra) over a number of repeat injections. Many reported resolution and mass accuracy numbers that you see are not RMS error numbers but instead originate from a single selected (favorable) ion.

It is important to remember in all applications that a weak signal (excessively high resolution) can yield poor ion statistics and can therefore be unusable. Too strong a signal can be equally useless, causing detector saturation. Ideally balanced ion statistics with definition in the spectra is the goal.

Some comparisons with respect to the figure:

  • quadrupole resolution is not sufficient to differentiate the two compounds in the upper figure.
  • with a resolving power of ~5000, TOF data clearly shows two distinct peaks, which can be accurately mass measured to < 5 ppm.

It is important to appreciate the various inter-related roles played in accurate-mass precision by the shift between definitions of mass and increasing resolution and factors such as peak shape and the need for calibration. If these are not clearly understood and taken into consideration mass mis-assignments and other undesirable results may occur.

The overlay figure shows quadrupole and TOF response where both mass values on the TOF data are within 1mDa of the exact mass.

The two fragments of different compositions in the figure are from the same analyte and therefore in the source at the same time. Even the best chromatography won't help in this case so this emphasizes one of the reasons higher resolution is useful especially in the analysis of unknowns. This applies equally well to QTof product ion data versus product ion data from a triple quad. As an added advantage with this higher degree of resolution the extracted ion current (XIC) plot of each allows differentiation of the oxygen containing and alkyl containing analytes selectively from the chromatograms where the quad data would lack this capability.


See MS – The Practical Art, LCGC 

  • Debating Resolution and Mass Accuracy, Vol. 22 No. 2, 118 – 130, February 2004
    • Why this is important: Deals with many of the issues at the heart of practical aspects of use as well as comparisons from industry sources
  • Petroleomics: MS from the ocean floor, Vol. 26, No. 3 March 2008
    • Why this is important: Discusses the utility of the highest resolution instruments and compares them to realistic expectations.

Also:

  • Methodology for Accurate Mass Measurement of Small Molecules, K. Webb, T. Bristow, M. Sargent, B. Stein, Department of Trade and Industry’s VIMMS Program within the UK National Measurement System, (LGC Limited, Teddington, UK 2004). VIMMS 2004 guidance document
    • Why this is important: A brief and concise overview of issues critical to success with measured, accurate-mass work.
  • Dealing with the Masses: A Tutorial on Accurate Masses, Mass Uncertainties, and Mass Defects, A. D. Leslie and D. A Volmer, Spectroscopy, Vol. 22, No. 6 June 2007
    • Why this is important: Expands on the basic topic to include Kendrick mass defect and labeling for peptides and proteins.

Terminology

Terminology

Sulphamethazine Nominal = 278

[C12H14N4O2S]

Average mass  Calculated using all isotopes of each element and their natural abundance.

Sulphamethazine avg. mass = 278.3313

[C12H14N4O2S]

Calculated exact mass  (Monoisotopic). Determined by summing the masses of the individual isotopes for a given ion.

Sulphamethazine exact mass = 278.0837

[C12H14N4O2S]

Accurate mass  (Actually “measured exact mass”). What we do with our instruments. It is the measure of an m/z reported to (typically) three or four decimal places.

As mass increases, differences between the definitions increase, and peak shape plays a bigger role:

Ubiquitin Nominal Exact Average

[C378H630N105O118S] 8556 8560.6254 8565.8730

Unit 16.1, An Overview of Peptide and Protein Analysis by Mass Spectrometry, S. Carr and R. Annan, in Current Protocols in Protein Science, J. Wiley and Sons (1996).

Related

Waters innovative single quadrupole mass detectors are designed for robustness, to make access to mass spectral data easy for scientists with all levels of expertise.

Waters innovative tandem quadrupole MS technology is designed for quantitative UPLC-MS/MS. Our philosophy centers on making high performance accessible, robust, and reliable.

Waters Time-of-flight MS systems, featuring QuanTof and MSE technologies, provide the highest UPLC-MS performance to meet the needs of challenging qualitative and quantitative applications.

Waters Quadrupole Time-of-flight MS systems, featuring QuanTof, Fast DDA and MSE technologies, provide the highest UPLC-MS/MS performance to meet the needs of challenging qualitative and quantitative applications.
Back To Top Back To Top