Mass spectrometers can be smaller than a coin, or they can fill very large rooms. Although the various instrument types serve in vastly different applications, they nevertheless share certain operating fundamentals. The unit of measure has become the Dalton (Da) displacing other terms such as amu. 1 Da = 1/12 of the mass of a single atom of the isotope of carbon 12 (12C).
Once employed strictly as qualitative devices-adjuncts in determining compound identity-mass spectrometers were once considered incapable of rigorous quantitation. But in more recent times, they have proved themselves as both qualitative and quantitative instruments.
A mass spectrometer can measure the mass of a molecule only after it converts the molecule to a gas-phase ion. To do so, it imparts an electrical charge to molecules and converts the resultant flux of electrically charged ions into a proportional electrical current that a data system then reads. The data system converts the current to digital information, displaying it as a mass spectrum.
Ions can be created in a number of ways suited to the target analyte in question:
The ions are separated, detected and measured according to their mass-to-charge ratios (m/z). Relative ion current (signal) is plotted versus m/z producing a mass spectrum. Small molecules typically exhibit only a single charge: the m/z is therefore some mass (m) over 1. The ‘1' being a proton added in the ionization process [represented M+H+ or M-H- if formed by the loss of a proton] or if the ion is formed by loss of an electron it is represented as the radical cation [M+.]. The accuracy of a mass spectrometer or how well it can measure the actual true mass may vary as will be seen in later sections of this primer.
Larger molecules capture charges in more than one location within their structure. Small peptides typically may have two charges (M+2H+) while very large molecules have numerous sites, allowing simple algorithms to deduce the mass of the ion represented in the spectrum.
Desorption methods (described in this primer) have extended the ability to analyze large, nonvolatile, fragile molecules. Routine detection of 40,000 Da within 0.01% accuracy (or within 4 Da) allows the determination of minor changes such as post-translational modifications. Multiple charging extends the range of the mass spectrometer well beyond its designed upper limit to include masses of 1,000,000 Da or more.
Natural isotope abundance is well-characterized. Though often thought to be stable, it can nevertheless display significant and characteristic variances. Isotope ratio measurements are used in metabolic studies (isotope-enriched elements serve as tracers) and also in climatic studies that measure temperature-dependent oxygen and carbon changes. In practice, complex molecules are reduced to simple molecular components before being measured using high-accuracy capabilities such as those found on magnetic sector instruments (see the following section).
Elemental analysis is typically performed on inorganic materials-to determine elemental makeup, not structure-in some cases using solid metal samples. Inductively coupled plasma (ICP) sources are common where a discharge (or lower power-glow discharge) device ionizes the sample. Detection using dedicated instruments, at the parts-per-trillion level, is not uncommon.