Scouting Gradients

Scouting Gradients

Although complex separations are most impacted by the selectivity of the column packing material, it is most efficient when developing a separation method targeted for purification to begin with optimization of the sample separation generated with a rapid scouting gradient. The goal of the scouting gradients is to determine the approximate solvent concentration required to elute the compound(s) of interest from the column. Scouting gradients typically start from the same strong solvent concentration (2–10%) and linearly increase to 25%, 75%, 50%, and 90–95%.

Run

Gradient start % B

Gradient end % B

Rapid gradient 1

2–10%

90–95%

Rapid gradient 2

2–10%

75%

Rapid gradient 3

2–10%

50%

Rapid gradient 4

2–10%

25%

Table 1. Example of rapid linear gradient scouting study.

Figure 4. Rapid scouting gradients. Resolution becomes higher as the gradient slope decreases.

Upon completion of the scouting gradients, the results are reviewed and evaluated by a series of questions before moving to the next step:

  • Does the rapid scouting gradient provide adequate separation for the compound of interest?
  • Is the rapid scouting gradient adequate for speed or efficiency?
  • Can the separation of the compound of interest from other peaks be maintained at the sample load required?

If the scouting gradient does not yield an adequate result, it can be optimized through method focusing, or as an alternative, the separation can be modified entirely through pH, solvent or other chromatographic variables.

Optimizing the Method

The scouting run that provides the greatest resolution for the compound of interest can be optimized by modifying the composition of strong and weak solvents. The composition can either remain constant (isocratic), or it can change per a given unit time or column volume (gradient).

Isocratic Elution

In an isocratic separation, weak solvent and strong solvent are maintained in an unchanging ratio per unit time. Isocratic solvent can be prepared offline by the analyst, or mixed online with a HPLC pump capable of proportioning the different solvents at a consistent, predetermined rate. This elution mode is convenient, consistent and robust because retention time influences caused by dwell volume are negligible when transferring a method between systems.

Isocratic elution has drawbacks and may not be appropriate for all separations. There are inherent issues including poor resolution of early eluting peaks, decreased symmetry due to peak tailing, decreased sensitivity due to band broadening, and problems with column contamination due to the buildup of strongly retained compounds.

Gradient Elution

Gradient separations in reversed-phase or ion exchange usually involve on-line mixing of mobile phase to achieve a steady increase in the organic solvent over the course of the analysis. At the beginning of the gradient when the solvent strength is low the analyte is partitioned in the stationary phase or remains at the head of the column. As the solvent strength increases, the analyte is transferred to the mobile phase, moves along the column, and finally is eluted.

Gradient elution provides the ability to separate analytes with a wide range of hydrophobicity in a reasonable time frame. Additional advantages can include improved peak resolution and increased sensitivity due to greater peak height. Column deterioration is also minimized when strongly retained components are washed from the column at the end of the run.

There are also disadvantages when using gradient elution. Consistent, pulse free online gradient mixing is required therefore expensive pump instrumentation is often necessary. Additionally, precipitation can occur when some mobile phases are mixed in certain combinations. Run times can be long due to post gradient re-equilibration, and instruments can vary in dwell time (Vd) causing method transfer issues, without adequate compensation.

Linear Gradient

Linear gradients progress with a constant increase in the composition of strong solvent throughout the entire run. They are often used in rapid screening experiments to quickly determine the solvent ratio that elutes the peak of interest and keep run time reasonably short, which results in elution in less time with less time and reduced solvent cost.

Figure 5. Solvent profile for a linear gradient.

Segmented Gradient

A segmented gradient is a combination of a series of linear gradients, each with a different slope or steepness. The shallow segments resolve the compounds of interest, while the steep segments are areas of the chromatogram where high resolution is not required, such as in a post separation wash.

Figure 6. Solvent profile for a segmented gradient.

Focused gradients are typically shorter and more exaggerated than in segmented gradients. Generally, the solvent strength is very low immediately after injection. It is then rapidly increased to between 2–5% below the solvent ratio required for elution of the compound of interest (i.e. 22% elution concentration – 2% = 20% start shallow gradient). A shallow gradient then proceeds at about 1/5 the slope determined in the rapid scouting separation to finally end between 2–5% above the elution concentration for the compound of interest (i.e. 22% + 2% = 24% stop shallow gradient). Finally, the percentage of solvent is quickly increased to wash the column of any remaining sample components.

Developing a Focused Gradient

The following calculations can be used to generate a focused gradient for the compound of interest after the sample has been analyzed with a rapid scouting gradient. Each equation is followed by a theoretical example.

First, the system dwell volume must be determined for the instrument that was used for the scouting gradient run. Instructions for dwell volume determination can be found in the section of this primer titled "Dwell Volume Determination" or it can be found in the “Analytical to Prep Gradient Calculator” application tool at www.waters.com/prepcalculator.

Additionally, column volume must also be determined for the column employed for the scouting run. It can be calculated by using the volume of a cylinder V = πr²h and the compensating for the volume occupied by packing material i.e. V = πr²h (66%). The “Analytical to Prep Gradient Calculator” application tool can also be used to perform this calculation.

Equation 1: Column volume

Column volume = π r² x Height x 66% available to mobile phase/1000 to convert mm³ to mL

Example:

CV = 3.14 x (4.6 mm/2)² x 50 mm x 0.66/1000
CV = 548 mm³ = 0.548 mL

Equation 2: Offset volume between gradient formation and the detector

Offset volume = System volume* + Column volume
*System volume or Dwell volume

Example:

Offset volume = 1.04 mL + 0.548 mL
Offset volume = 1.588 mL

Equation 3: Time to detector

Time to detector = Offset in mL/Flow rate in mL/min

Example:

Time to detector = 1.588 mL/1.5 mL/min
Time to detector = 1.06 min 

Equation 4: Time when elution concentration was formed

Time elution of concentration = Peak of interest retention time – Time to detector – Gradient hold

Example:

Time of elution concentration = 6.0 min – 1.06 min – 0.0 min
Time of elution concentration = 4.94 min

Equation 5: Percent elution concentration

% Elution concentration = Time elution concentration/Length scouting gradient segment x Change scouting + Initial scouting gradient %

Example:

% Elution concentration = 4.94 min/6.0 min x 90% + 5%
% Elution concentration = 79.1%

Equation 6: Number of column volumes (CV)

# Column volumes = 1 column volume/mL x Flow mL/min x length scouting gradient segment

Example:

# CV = 1 CV/0.548 mL x 1.5 mL/min x 6 min
# CV = 16.4

Equation 7: Scouting gradient slope

Scouting gradient slope = % change scouting gradient/# column volumes

Example:

Scouting gradient slope = (95% – 5%)/16.4 CV
Scouting gradient slope = 5.49%/CV

Equation 8: Focused gradient slope

Focused gradient slope = 1/5 x scouting gradient slope

Example:

Focused gradient slope = 1/5 x 5.49%/CV
Focused gradient slope = 1.1%/CV

The focused gradient segment is created from 5% below, to 3–5% above the estimated elution percentage. As an example, the elution percentage is 79%.

Equation 9: Length of time for the focused gradient segment

Length of time for the focused gradient segment = % range x (1/scouting gradient slope) x column volume x (1/ flow rate)

Example:

Length of time for the focused gradient segment = (79%+5%) - (79%-5%) x (1/1.1%) x (0.548 mL) x (1/1.5 mL/min)
Length of time focused gradient segment = 3.32 min

The last step is to write the focused gradient using the values calculated in the previous formulas.

Scouting gradient:

Time (mins)

Flow rate (mL/min)

Solvent

%A

%B

0

1.5

95

5

7

1.5

5

95

8

1.5

5

95

9

1.5

95

5

Focused gradient:

Time (mins)

Flow rate (mL/min)

Solvent

%A

%B

0

1.5

95

5

1

1.5

26

74

4.32

1.5

16

84

6

1.5

5

95

7

1.5

95

5

Related

Purify micrograms to milligrams of your compounds with Waters flexible analytical to semi-preparative purification solutions.

Get the flexibility of high-throughput parallel runs for selective mass-directed fraction collection with hundreds of samples with Waters AutoPurification HPLC-MS System.

Increase throughput with fully automated UV- or mass-directed semi preparative to preparative-scale SFC purification systems from Waters.

Achieve fast, efficient, lab-scale separations, and greater throughput with Waters Preparative HPLC Columns, which provide predictable performance and easily scale from analytical to prep.
Back To Top Back To Top