The room temperature GPC applications for organic soluble polymers were all performed on the Alliance System. In most cases, the Styragel HR columns were used for the analysis. Because of the unique design of the solvent manager in the Alliance system, the flow rate precision is better than 0.075%, and the flow is virtually pulseless. Pulseless flow is extremely important for people doing light scattering, as a pump that pulses will loosen the column "fines" and cause the spikes in the chromatogram.
The first step in any GPC analysis is to calibrate the system. Below, you see a polystyrene narrow standard calibration curve that was obtained on Alliance® using THF as the eluent. The molecular range covered is ~250 to 3M. The column set consisted of 2 HR 5E's (mixed bed) and a single HR2 (500 Å). The columns were heated to 40 °C in the column heater, and the flow rate was 1.0 mL/min. The calibration curve is a 5th order fit. The curve looks excellent, but there is also something very interesting to note. There are three injections of each standard shown on the curve, (3 injections each from three different vials). So the total number of points on the curve is 39! (If you look hard enough you may be able to see some evidence of a very small amount of scatter for a couple of the standards). The retention time reproducibility of the narrow standards is less than 0.04%, a result of the superior flow delivery of the Alliance System. We sometimes need to do our GPC analysis in solvents other than THF.
Below is a narrow standard calibration curve using poly(methyl methacrylate) standards, with dimethylformamide being used as the eluent.
We prefer to use PMMA's rather than polystyrene when working with DMF, as low molecular weight polystyrene standards tend to have inconsistent retention times, eluting later than expected. The polystyrene oligomer standards, (molecular weight under ~700, for example), may show retention beyond the total volume, VT. The PMMA narrow standards do not exhibit this tendency, and are preferred for work in DMF. The same column set that was used for the polystyrene calibration was used here, (2 HR 5E's plus a single HR2). The only difference is that these columns were packed in DMF. Lithium bromide, at a concentration of 0.05M, was added to the DMF. This is to prevent any polar interaction between sample and eluent, as most samples run in DMF tend to be very polar. As for the polystyrene curve, there are three injections for each standard, so, in this case, there are 36 calibration points on the curve. There seems to be even less scatter on this curve than the polystyrene curve. The columns were heated to 80 °C to reduce the viscosity of the DMF.
A well characterized broad polystyrene standard, Dow 1683, was run on the Alliance system with the dRI detector and THF as the eluent. The concentration was 0.15%, and a 300 μL injection was made. The broad standard was injected again, (also 300 μL) but this time at a 0.015% concentration (10 times less). You can see the comparisons in the figures below. Note that the signal for the 0.15% concentration is ~15mV. With the baseline noise being 14 uV, we have a S/N of >1000:1. The signal for the 0.015% injection is only 1.5 mV, but we still are able to attain a S/N of >100:1, and is easily integrated. The smoothness of flow, along with the integrated degasser, allows us to run very dilute polymer solutions, and still get the S/N needed for reproducible GPC work. This is very important when we need to run very low concentrations of high molecular weight samples. We can now run extremely low concentrations and still get the correct results without sacrifice of S/N.
Determining the molecular weight distribution of elastomers (both natural and synthetic) is a very important analytical technique used to correlate with physical properties. Elastomer formulations may be very complicated, with blends of polymers being used, as well as antioxidants, plasticizers, vulcanizers, accelerators, and a variety of fillers (carbon black, titanium dioxide, silica, etc.). The entire formulation may consist of only 50% (or even less) of the elastomer. These formulations are used extensively in the automotive and aerospace industries for everything from tires to O-ring seals. As is always the case in GPC analysis, the first thing we must do is calibrate our system, so here we show athird-order calibration curve using polybutadiene narrow standards as the calibrants.
There are also polyisoprene narrow standards available. Once again, two HR 5E's and a single HR2 were used for the column bank, maintained at 75 °C. In the case of elastomers, toluene is usually the solvent choice. THF may be used in many cases, but toluene tends to do a better job at dissolving some elastomers such as natural rubber, (cis - 1,4 polyisoprene). The dRI detector was used with the Alliance system. We chose polybutadiene narrow standards as they are similar in structure to most of the elastomers we looked at. Note the outstanding reproducibility for the applications that show the multiple distribution overlays.
Below are a few additional elastomer applications of interest:
The GPC analysis is pretty straightforward, using either THF or methylene chloride as the eluent. We decided to see how good the precision was for the Alliance System by doing something a little different. We had a series of polycarbonates that we ran GPC analysis on Alliance System in Milford, and had the same samples run on an Alliance System at a Waters site outside the U.S., with even a slightly different column set. Shown below are the amazing agreement obtained between the two lab sites.
Aqueous GPC analysis brings a whole new set of challenges to the polymer characterization chemist. Most conventional, high performance packings for aqueous GPC analysis are prepared from hydrophilic methacrylate gels, with residual carboxylate groups, giving the column chemistry an overall anionic charge. When doing GPC analysis on water soluble polymers, one must be cognizant of the fact that there could be a charge interaction between the sample and the packing material, unless certain steps are taken. Theoretically, if the polymer is neutral, you could do the analysis in pure water. If there is any anionic charge to the polymer, it will be excluded by the column and elute at the void volume if pure water is used as the eluent. On the other hand, if the polymer has an overall cationic charge, (and you use pure water as the eluent), the sample will stick to the column and never elute. A lot of these ionic problems can be overcome quite easily with the addition of an electrolyte, such as 0.10M NaNO3.. Even for neutral samples, it is a good idea to use 0.10M NaNO3 as the eluent. Some of the problems that need to be overcome with the correct eluent (see solvent selection guide for water soluble polymers) are as follows:
There are occasionally other interactions that can occur, such as association effects and memory effects, but the 5 problems shown above are the ones most encountered. The aqueous solvent selection guide shown previously will help you to choose the correct eluent for you particular application. We have chromatograms for nearly all of the applications shown in the guide. Just let us know if you need any help with your particular samples.
Three different water soluble polymers were run on the Alliance system, with refractive index detection. The three polymers analyzed (shown below) were Hydroxyethyl cellulose, Pectin, and Polyalginic Acid. Note the excellent reproducibility of the multiple molecular weight distribution overlays. In all cases a three column Ultrahydrogel set (2 Linear plus a 120) was used. Note that the eluent was 0.10M sodium nitrate, an excellent choice for neutral and anionic hydrophilic polymers.
The GPC analysis of nylons and polyesters have historically been very difficult to do, with m-cresol at 100 °C being the solvent choice for many years. There have been a variety of other solvent choices that workers have tried, but one that works quite well is hexafluoroisopropanol, (HFIP). HFIP has an advantage over m-cresol in that nylons and polyesters dissolve at room temperature. One disadvantage is cost: HFIP costs approximately $1,000 per liter. This is the reason we have investigated the GPC analysis of these two popular polymers using HFIP with solvent efficient columns, which are 4.6 mm i.d. These are 30 cm long columns, but the more narrow i.d. (as compared to conventional 7.8 mm columns) allows for a large savings in solvent use (and disposal costs). The flow rate is usually ~0.35 mL/minute, which will give approximately the same eluent linear velocity as 1.0 mL/minute with the 7.8 x 300 mm columns we usually use. These solvent efficient HFIP columns are specially packed in methanol for conversion directly to HFIP at 0.05 mL/minute.
For our analysis of nylons and polyesters, 0.05 M sodium trifluoroacetic acid was added to the HFIP, to prevent any polar interactions. Nylons in particular will exhibit tailing on the low molecular weight end if the salt is not added to the HFIP. Once again, the Alliance GPC System with dRI detector was used for the analysis. Because of the low system volume (low dispersity) of the Alliance System, excellent resolution may still be obtained with the 4.6 mm columns. We used a column designation of HR2, HR3, and HR4, which represents high resolution columns in the 500, 103, and 104 angstrom range. The RI and the columns were maintained at 30 °C and the injection volume was only 25 µL for the narrow PMMA standards and samples. Polystyrene does not dissolve in HFIP, so the narrow poly(methyl methacrylate) standards are used, and they work very well.
Here we show a third-order calibration curve for the PMMA standards in HFIP (triplicate injections of each standard). The extraordinary retention time reproducibility of the standards is obvious from the curve.
The first set of samples run was Poly(ethylene terephthalate), (PET) and Poly(butyleneterephthalate), (PBT) shown here.
Also shown below is an overlay of 5 molecular weight distributions of Nylon 6/6. A Nylon 6/6 broad standard was used for the calibration, so the molecular weights shown are "accurate" for the nylon sample.
The last work shown in HFIP is for two medical plastic grade Polyether/amide copolymers, used to make catheters. The two samples did not have the same physical properties nor the same "processability", yet FTIR, thermal analysis, rheological measurements, melt flow index, etc., showed no discernible differences between the two samples.
If you look at the two molecular weight distributions individually (shown here as 5 overlays), they indeed do look quite identical to eachother.
However, if you look at the overlays of the 5 MWD's for each here, you can easily see some differences between the two. The reproducibility of the Alliance System gives you the confidence to say that these small MWD differences are indeed real, and not due to injection-to-injection variability.
In recent years there has been increased interest in using gradient HPLC techniques, such as Gradient Polymer Elution Chromatography (GPEC), with polymers for determining the compositional drift of copolymers, the composition of polymer blends, or for the analysis of polymer additives. Depending upon the gradient conditions and columns selected for analysis, separations may be obtained dependent on molecular weight or based upon precipitation, or adsorption mechanisms. The use of an Evaporative Light Scattering Detector (ELSD) allows one to perform solvent gradients with a universal mass detector and observe both UV absorbing and non-UV absorbing polymer samples without baseline disturbances from the solvent gradient. The addition of a Photodiode Array Detector (PDA) allows for compositional analysis across the molecular weight distribution of many copolymers, can be useful for the identification of components in a polymer blend, and also is invaluable for the quantitation of polymer additives and other small molecules in traditionalreverse-phase separations.
This section demonstrates the advantages of gradient analysis of polymers as compared to results obtainable with Gel Permeation Chromatography. The instrumentation used to carry out this work is described and examples of this technique for the analysis of polymer blends are shown. The effects of column functionality and solvent composition on the separation of polystyrene standards and samples is described and the best conditions observed are used to analyze various copolymers for monomer composition. Finally, the traditional use of gradient separations with the same instrumentation for the analysis of several types of polymer additives is also shown.
The most common chromatographic method for the analysis of polymers is Gel Permeation Chromatography (GPC) where the separation is based upon the size of the polymer sample in solution, or the hydrodynamic volume of the polymer solution.Figure 1 shows the chromatograms obtained using GPC for a polystyrene sample, polystyrene-acrylonitrile copolymer (25% acrylonitrile) and a polystyrene-butadiene rubber (50% styrene) analyzed separately.Even though the samples are of different molecular weight, the hydrodynamic volumes are similar enough that the polymer peaks are observed at nearly the same retention time.The chromatograms obtained for the GPC analysis of a blend of approximately the same concentration of each of the polystyrene, polystyrene-acrylonitrile, and the polystyrene-butadiene samples are also shown in Figure 1. This chromatogram shows no separation of the three different polymers and demonstrates the impracticality of GPC for the analysis of most polymer blends.
However, when this same polymer blend is analyzed in a gradient mode, the three components can easily be baseline resolved as demonstrated in Figure 2 which shows the overlay of two replicate injections of the polymer blend run on a prototype divinylbenzene-vinylpyrolidone column with a gradient from 100% Acetonitrile (ACN) to 100% Tetrahydrofuran (THF) over 20 minutes.
Using this technique, the samples are dissolved in THF and then injected into the chromatographic system running 100% ACN. The polymers in the blend are insoluble in acetonitrile and precipitate onto the column. As the gradient proceeds, the polymers in the blend are redissolved according to their solubilities and are eluted from the column as well resolved peaks. This mechanism is similar to Gradient Polymer Elution Chromatography (GPEC). Other gradient methods for the analysis of polymers have been described in the literature which are performed under conditions where the polymers remain in solution and are separated by an adsorption mechanism, but these are generally for polar polymers that are soluble in alcohols or ketones run on bare silica columns and are not discussed here.
All gradient work was carried out using the following system configuration unless otherwise noted.
System: |
Waters Alliance 2690 Separations Module with column heater at 30 ºC |
Detector 1: |
Waters 996 Photodiode Array Detector |
Detector 2: |
Alltech Model 500 ELSD with LTA Adapter (Drift Tube at 40 ºC, 1.75 Liters/min Nitrogen) |
Data system: |
Waters Millennium 32 Chromatography Manager |
Column: |
As listed in Figures, 30 ºC |
Flow rate: |
1 mL/min |
Samples: |
10–25 µL injections of 0.2–0.5% samples |
Gradient: |
Linear gradient, conditions and mobile phases as listed in Figures. |
The most commonly used detector for GPC is the Refractive Index (RI) detector; however, the sensitivity of the RI to changes in mobile phase composition makes it unacceptable as a detector for Gradient Polymer Analysis. Figure 3 shows the chromatograms obtained for the 25 µl injection of a 0.5% solution of a styrene-acrylonitrile copolymer (25% Acrylonitrile) run on a prototype DVB/Vinylpyrolidone column with a gradient from 100% ACN to 100% THF in 20 minutes using a refractive index detector (RI), a photodiode array detector (PDA), and an evaporative light scattering detector (ELSD).
As soon as the mobile phase change from the gradient reaches the RI detector (~2.5 minutes) the RI signal goes offscale, completely overloading the detector. The chromatogram obtained from the PDA detector at 260nm (or any UV detector) demonstrates that UV detection is much better suited for gradient analysis than RI detection. The chromatogram does show baseline drift with the change in mobile phase but there is still good sensitivity for the polymer sample and the drift can easily be eliminated by baseline subtracting a blank gradient run. The third chromatogram in Figure 3, obtained using an ELSD, demonstrates the superior performance of the ELSD for gradient applications. The detector is essentially insensitive to the changes in the mobile phase composition since the solvents are evaporated prior to detection. This, combined with the excellent sensitivity for polymer samples, makes the ELSD the detector of choice for gradient analysis of polymers. By combining a PDA with the ELSD, one can detect and quantitate unknowns with the ELSD and use the PDA to determine peak purity, for the identification of unknowns through library matching, and for compositional analysis of copolymers.
Using this system, a wide variety of different types of polymers, polymer blends and copolymers can be analyzed. Figure 4 shows an overlay of chromatograms obtained for many types of polymers run on a Nova-Pak C18 Column with a 30 minute gradient from 100% ACN to 100% THF including polyvinylchloride, polymethylmethacrylate, polystyrene, polystyrene-butadiene block copolymer, polydimethylsiloxane, polystyrene-isoprene block copolymer, and butyl rubber.
When using this technique for the analysis of polymer blends or copolymers, it is necessary that the separation be independent of molecular weight so that the polymers are separated only by composition. Unfortunately, since this is primarily a precipitation/redissolution mechanism, some molecular weight dependence is inevitable, but it can be minimized through the judicious selection of columns, mobile phases, and gradient conditions.
Figure 5 shows an overlay of chromatograms obtained from a series of narrow polystyrene standards run on a SymmetryShield C8 Column (3.9 mm x 15 cm) with a gradient form 100% ACN to 100% THF in 10 min.
The standards from 43,900 to 2,890,000 MW elute in a band from approximately 9 to 9.5 minutes. The lower MW standards elute earlier, with many of the oligomers well resolved. These lower molecular weight standards are soluble or nearly soluble (9100 MW) in the starting conditions of the gradient (100% ACN) and are therefore separated by the traditional reverse-phase mechanism. Figure 6 shows an overlay of the chromatograms obtained for the same standards run under identical conditions on a prototype DVB/vinylpyrolidone column (3.9 mm x 15 cm).
A similar pattern is observed with the 43,900 to 2,890,000 standards eluting in a slightly narrower band. The separation of the lower molecular weight standards is somewhat different; however, this is not surprising due to the different reverse-phase characteristics of the two columns.
By changing to a Nova-Pak C18 Column (3.9 mm x 30 cm) and using a 30 min gradient, the chromatograms shown in Figure 7 were obtained. Using these conditions, the molecular weight dependence for the 43,900 MW and higher polystyrene standards is nearly eliminated. As expected the lower MW standards that are soluble in ACN are eluted earlier in the chromatogram, however, the low MW oligomers are being split into three peaks, indicating that they are being separated by their differing end groups.
The choice of mobile phase used as the non-solvent can have significant effects on the separations obtained from gradient analysis of polymers.
Figure 8 shows an overlay of chromatograms obtained for the same standards run on a Nova-Pak C18 Column (3.9 mm x 15 cm) with a linear gradient from 100% Methanol (MeOH) to 100% THF in 30 min. These results show a clear dependence on molecular weight from the well-resolved oligomers early in the chromatograms to the 8 million MW standard. This is undesirable for the purposes of copolymer or polymer blend analysis, as it would be difficult to determine whether differences in retention time were due to compositional differences or MW differences.
This non-solvent effect can also be seen when analyzing broad MW polymer samples.
Figure 9 shows the chromatograms obtained for NBS706 broad polystyrene standard run on a Nova-Pak C18 Column (3.9 mm x 15 cm) with a 30 min gradient using first ACN and then MeOH as the non-solvent with THF as the solvent for both injections. When using ACN as the non-solvent, a more desirable sharp peak is obtained whereas when MeOH is used as the non-solvent, a very broad peak is obtained. Our work has shown that for THF soluble polymers, the best separations were observed with the 100% ACN to 100% THF gradient. These conditions give a rugged method that can be used for a wide variety of polymer blends and copolymers.
Gradient analysis is a powerful tool for evaluating copolymer materials. A series of random styrene-butadiene rubbers (SBR) were run using this 100% ACN to 100% THF gradient on a prototype DVB/Vinylpyrolidone column (3.9 mm x 15 cm) in 20 min. Five different SBRs with composition ranging from 50% styrene to 5.2% styrene were injected along with a narrow polystyrene standard (355K MW) and a narrow polybutadiene standard (330K MW). An overlay of the resulting chromatograms is shown in Figure 10.
The different SBRs are easily separated by their relative amounts of styrene and butadiene. These SBRs were previously analyzed by traditional GPC to be sure that the molecular weights were high enough that molecular weight dependence would be insignificant, and the molecular weights were all found to be approximately 200,000 to 300,000 by relative calibration with polystyrene.
Using the gradient results, a calibration curve was constructed to determine % styrene vs retention time and is shown in Figure 11.
The plot exhibits a good correlation between % styrene and retention time so that this method could be used to determine the approximate composition of an unknown SBR. The UV data from the PDA could also be used to crosscheck the results from the ELSD.
In a similar manner, Figure 12 shows the chromatograms obtained for a series of block styrene-butadiene copolymers with a similar separation as the random SBRs.
The data is plotted in Figure 13 showing a calibration curve similar to the one obtained for the random SBRs. Using this gradient method, species with only slight differences in structure can easily be separated.
Figure 14 shows an overlay of individual injections of polymethylmethacrylate, polymethylmethacrylate, poly-n-butylmethacrylate, poly-n-hexylmethacrylate, and poly-laurelmethacrylate run on a Nova-Pak C18 Column (3.9 mm x 15 cm) with a gradient of 100% ACN to 100% THF in 30 minutes. The chromatograms show excellent separation between each component in the homologous series of methacrylates and could easily be resolved with a faster gradient.
The chromatogram in Figure 15 shows the separation of the same methacrylates injected as a mixture and run under identical conditions demonstrating an identical separation when the components are run in a mixture.
This same method using identical conditions also has utility for analyzing lower molecular weight compounds. Figure 16 shows an overlay of chromatograms for two low molecular weight waxes. The two waxes are well separated and slight differences between the oligomer ratios can be observed.
Low molecular weight polymer additives can be analyzed with this method by the traditional reverse-phase mechanism. Many types of polymer additives will be shown using the following conditions that were chosen to be compatible with a mass spectrometer:
System: |
Waters Alliance 2690 Separations Module with column heater at 30 ºC |
Detector 1: |
Waters 996 Photodiode Array Detector |
Detector 2: |
Alltech Model 500 ELSD with LTA Adapter (Drift Tube at 40 ºC, 1.75 Liters/min Nitrogen) |
Data system: |
Waters Millennium 32 Chromatography Manager |
Column: |
Symmetry C8 ,2.1mm x 15cm, 30 ºC |
Flow rate: |
0.29 mL/min |
Gradient: |
Linear Ternary Gradient, 30 mins; 70/10/20 to 1/79/20 H2O/ACN/THF |
Figure 17 shows the separation of Tinuvin 440, Tinuvin 900, and Tinuvin 328 that are UV stabilizers commonly used in polyolefin resins. Even though these compounds are difficult to extract from polyolefin resins with good recovery, once extracted they can be analyzed easily with good sensitivity using this method.
Several different types of phthalate plasticizers are separated in Figure 18. Phthalates, which are commonly used as plasticizers in PVC resin, have come under scrutiny recently as possible carcinogens. Phthalates, particularly diethylhexylphthalate (DEHP), are used routinely in medical devices such as catheters and IV bags and in children's toys possibly exposing patients and children to high levels of this suspected carcinogen. This method is a simple means for analyzing these phthalate compounds.
Figure 19 shows the chromatograms for the slip agents oleamide and erucamide and the antistat stearic acid. These compounds, which have very little UV absorbance, exhibit poor sensitivity with UV detection but can easily be detected with the Evaporative Light Scattering Detector.
Figure 20 shows the separation of Irganox 1076 and Irgafos 168 that are two antioxidants commonly used in polyolefins and other polymers. Irganox 1076 is a hindered amine and Irgafos 168 is a phosphite ester that degrades easily. The chromatogram shows two peaks for Irgafos 168. The second peak is the main Irgafos 168 peak while the first peak is actually the oxidized Irgafos 168 impurity that was present in the sample. This method is not meant to be an optimized method but only a general method for use with a wide variety of additives.
Figure 21 shows 12 overlays of a separation of 10 common antioxidants run using a modified version of the approved ASTM method for the analysis of additives in polyolefins. The column, mobile phases, flow rate, and gradient conditions were optimized to obtain the shortest analysis time and maximum sensitivity allowing for the analysis of these 10 antioxidants in less than 10 minutes.
The method utilizes both a mobile phase gradient and a flow rate gradient resulting in an extremely reproducible and sensitive method. The analytes were detected with a PDA at 230 nm which besides giving excellent sensitivity, also allows for peak identification using the library matching capabilities of the photodiode array detector. The instrument and conditions used to carry out this separation are shown in Figure 22.
The use of gradient methods for the analysis of polymers allows for separations that are essentially independent of molecular weight. Individual polymers in blends having the same molecular weight distribution can easily be separated and copolymers can be separated by their monomer ratios. Using the same instrumentation, mostcommon polymer additives may also be analyzed. The Evaporative Light Scattering Detector is a universal detector which is unaffected by changes in mobile phase gradient composition and the Photodiode Array Detector allows for positive identification of many compounds and compositional analysis of copolymers. These gradient methods are highly reproducible techniques and are extremely well suited for deformulation applications.
People involved in polymer characterization by chromatographic techniques do not exclusively use GPC to analyze their samples. Many times we need to use liquid chromatographic techniques by adsorption or partition chromatography to get the information we need.
Conventional reverse-phase and, at times, normal phase separation techniques are used to quantitate polymer additives, as an example. Obtaining the molecular weight distribution of your polymer sample may be just one part of the characterization process. What about the additives that are formulated into the polymer to offer stabilization or processing enhancement? They can be even more important than the polymer itself. We need to think about using the correct UV stabilizers and antioxidants for protection against degradation, plasticizers to improve flexibility, antistats for polyolefins, flame retardants, accelerators to enhance the crosslinking (or curing) process, and so forth.
We have done an extensive amount of work with polymer additives, and you can find some of our published work detailed in the Journal of Liquid Chromatography, volume 14 #3, (1991) and volume 16, #7, (1993).
How do we analyze polymer additives? First, we need to think of what we are trying to accomplish. Do we need to know if the correct amounts of each additive are present in the formulation? Are we trying to "deformulate" a competitive material? Do we need to extract the additive package out from the polymer matrix? Chances are, the answers to these questions will be "Yes". GPC analysis is not the best way to separate, identify and quantitate the levels of additives present. Most of the additives are quite close to each other in size and molecular weight, so we need to use HPLC to separate them. A simple gradient technique, with optional flow programming, works very well in getting many different types of additives separated in a short run time. A gradient analysis consists of varying the eluent, or mobile phase composition, usually from a "weak" solvent to a "strong" solvent over a period of time. This composition variation is usually done in a linear fashion for additive analysis. Since we are varying the eluent composition throughout the chromatographic run, the refractive index detector can not be used.
Most of the polymer additives we deal with have some chromophore that absorbs ultraviolet light, so a UV detector is used primarily. If there are no chromophores present, an evaporative light scattering detector may be used. We can also change the flow rate throughout the run, usually increasing flow to get the later eluters to come out more quickly. The column usually chosen for additive analysis is either an octadecylsilane (C18) or Octylsilane (C8) column, ~15 cm in length. An example of a reverse-phase gradient (with flow program) separation of a series of common antioxidants and UV stabilizers 9 overlay of 12 injections is shown here.
The gradient conditions are quite simple: 70% acetonitrile/30% water initially, then proceed to 100% acetonitrile in a linear fashion after just 5 min. There is also a flow program, from 2.0 mL/min initially to 6 min, then ramping to 3.0 mL/min in just 12 sec. The table of data shows the remarkable reproducibility results (retention time and area RSD's) for each additive. This is further testimony to the incredible flow and sample delivery reproducibility of the Alliance System.
The UV detection was carried out at 230 nm. The PDA detector looks at all wavelengths (that you choose to view) simultaneously, which allows you to obtain the UV spectra for each additive. This spectrum may then be stored in a library and compared to a stored library of known additive standards. The only drawback to the library search is that a large majority of antioxidants are hindered phenols, which all have very similar spectra. In this case, you will have to rely only on the retention time for identification purposes. Another alternative is to add a mass spectrometer detector to the system. This will provide an electron impact spectrum which is library searchable.
Beginner's Guide to Size-Exclusion Chromatography
Frequently Asked GPC & SEC Questions
Applications for Room Temperature GPC