Derivitization of Amino Acids Using Waters AccQ•Tag Chemistry

Derivitization of Amino Acids Using Waters AccQ•Tag Chemistry

6.1 INTRODUCTION

6.1 INTRODUCTION

Amino acid analysis by liquid chromatography and optical detection requires additional sample preparation, because most amino acids lack a chromophore and cannot be detected. Thus, after hydrolysis, the next step in an amino acid analysis is derivatization. This section describes the preparation of protein or peptide hydrolysates for derivatization using Waters AccQ•Tag chemistry.

To properly and reliably derivatize a hydrolyzed sample, the following must be considered:

  • Removal of particulates
  • Determination of sample mass for derivatization
  • Determination of whether sample neutralization is required
  • Use of excess derivatization reagent based on sample mass

The following discussions are abbreviated out of necessity. For additional information and guidance on the AccQ•Tag derivatization chemistry, please visit www.waters.com/AAA.

6.2 REMOVE PARTICULATE MATTER

6.2 REMOVE PARTICULATE MATTER

Centrifugation may be necessary, if large amounts of particulates or floating lipids are present. Centrifugation makes it is easier to withdraw an aliquot of clear hydrolysate for derivatization.

For large volume samples such as feed analysis hydrolysates, filtration of the particulates may be enough, noting that sample recoveries may be affected by choice of filter material. This factor needs to be considered and addressed in order to obtain unbiased results. Contact the filter manufacturer for details regarding filter performance in this application.

6.3 DERIVATIZATION USING WATERS AccQ•Tag CHEMISTRY

6.3 DERIVATIZATION USING WATERS AccQ•Tag CHEMISTRY

The AccQ•Tag Method is a pre-column derivatization technique for amino acid analysis of hydrolyzed peptide and proteins. The AccQ•Tag Method accomplishes the following:

  • Uses Waters AccQ•Tag Ultra or AccQ•Fluor Reagent to derivatize the amino acids
  • Separates the derivatives using gradient-based, reversed-phase HPLC or UPLC
  • Accurately quantitates the derivatives using external amino acid standards and optical detection

The 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate (AQC) reagent reacts with both primary and secondary amines. Excess AQC reagent reacts with water to form 6-aminoquinoline (AMQ). AMQ reacts slowly with excess AQC reagent to form a bis urea. These side products do not interfere with the separation, identification, and quantitation of the amino acids contained in the peptide or protein hydrolysates. The derivatives are stable for days, permitting batch processing or repeat analysis, if desired.

Figure 10. Schematic of AccQ•Tag reaction from derivatization of primary and secondary amino acids.

Over many years, extensive studies have been performed to ensure the accuracy of the AccQ•Tag derivation reaction. The chemical reaction itself requires both a molar excess of the derivatization reagent and basic pH (8–10) for complete derivatization of all the amino acids. Strategies have been developed to address these critical factors.

6.3.1 Determination of sample amount for AccQ•Tag derivatization

6.3.1 Determination of sample amount for AccQ•Tag derivatization

In the reaction itself, the sample is typically diluted 10x, and 1 µL out of a 100 µL total derivatization volume is injected onto the column. Because not all amino acids are present in the same amount in a derivatization, the amount of sample must be sufficient for the least abundant amino acid to impact the limit of detection or quantitation. Derivatization of at least 1 pmol of the least abundant amino acid, in a maximum injection volume of 1 µL, is recommended to obtain accurate quantitation. To determine the amount of sample needed, perform the following calculation:

Example calculation:
For a sample concentration of 1.0 mg/mL protein:
Estimate the amount of sample needed for the least abundant amino acid. In this example, we assume 0.03 mg/mL is needed to deliver 1 pmol on column of this amino acid.

Step 1: Determine approximate molar concentration of the least abundant amino acid.

Step 1: Determine approximate molar concentration of the least abundant amino acid.

Convert mg to moles (average molecular weight of an amino acid residue in a protein is 110).

This is the estimated concentration of the least abundant amino acid in this sample.

Step 2: Determine the dilution required to give 1 pmol of this least abundant amino acid on column.

Step 2: Determine the dilution required to give 1 pmol of this least abundant amino acid on column.

This gives a 27-fold dilution (10 ÷ 0.37 = 27).

The hydrolysate requires a 27-fold dilution prior to derivatization. For example: 5 µL of the sample hydrolysate can be diluted by 135 µL of 0.1 N HCl to yield this target value.

Finally, a 10 µL aliquot of the above dilution can be transferred to the derivatization vial.

6.3.2 Neutralization for AccQ•Tag derivatization

6.3.2 Neutralization for AccQ•Tag derivatization

To ensure complete AccQ•Tag reagent derivatization of the amino acids contained in the hydrolysate, the sample must be buffered in an approximate pH range of 8.2 to 10.1. If the acid hydrolysate is not properly neutralized, and if the pH falls below 8.2, derivatization will be incomplete. The effect of pH varies for each amino acid. Note that not all of the amino acids are affected equally. Acidic amino acids, such as glutamic acid and alanine, are impacted more than serine or phenylalanine. pH is a critical factor in obtaining accurate quantitation of all the amino acids in the original sample.

  • If the amino acid solution is dissolved in 0.1 N HCl, 10–20 µL of sample can be transferred directly to the derivatization cocktail without pH adjustment. Note: It is still necessary to ensure that the available derivatization reagent is in excess, as described below.
  • If the amino acid solution is in a higher concentration of acid (>0.1 N HCl), it should be neutralized with an equal volume of sodium hydroxide at the same concentration. This can either be done as a bulk volume addition or integrated into the derivatization step.
    • Substitute the required amount of borate buffer with NaOH to neutralize the HCl in the sample.
    • In the derivatization reaction vial mix, place 10 µL of xM NaOH and 60 µL of borate. Add 10 µL of the AA sample (in xN HCl). Derivatize with 20 µL of AccQ•Fluor reagent.
  • If there is a question about proper neutralization, you can prepare trial samples and check the final pH with single-use pH strips.

WARNING: If the sample turns bright yellow upon addition of the derivatization reagent, the pH of the sample is too low. Neutralize with NaOH.

Example calculation:
To determine the amount of NaOH needed for neutralization, perform the following calculation.
For the above sample of protein at 1.0 mg/mL in 6 N HCl, which needs to be diluted 27x to ensure that there is enough excess derivatization reagent in the sample, the following calculations apply.

Step 1: Calculate the final acid concentration of the sample after dilution.

Step 1: Calculate the final acid concentration of the sample after dilution.

Convert the acid concentration of the sample from molar to µmoles:

Determine the final concentration of acid in the diluted sample:

Reminder: 5 µL of the sample hydrolysate can be diluted by 135 µL of 0.1 N HCl to yield this target value.

Step 2: Determine the amount of base (NaOH) to add to the buffer for derivatization.

Step 2: Determine the amount of base (NaOH) to add to the buffer for derivatization.

The total amount of NaOH needed per derivatization is 0.31 M.

Since the total borate buffer added to derivatization is 70 µL, there are two methods for neutralization:

Add 10 µL of 0.31 M NaOH and 60 µL of buffer for each derivatization.

In a separate vial, mix 600 µL of borate buffer and 100 µL of 0.31 M NaOH. Mix. Add 70 µL of this mixture + 10 µL sample + 20 µL AccQ•Tag derivatization reagent for each sample.

6.3.3 Derivatization reagent excess confirmation for AccQ•Tag derivatization

6.3.3 Derivatization reagent excess confirmation for AccQ•Tag derivatization

For complete derivatization of all the amino acids, 4–6x molar excess of the AccQ•Tag derivatization reagent is needed in the reaction. If there is insufficient reagent, some relatively sensitive amino acids will not be completely derivatized. The rates of derivatization for each amino acid vary based on the chemical properties of the amino acids; for example, the recovery of alanine can be significantly affected by insufficient molar excess of AccQ•Tag, while phenylalanine is more immune to these effects.

To determine the amount of sample to add to the reagent vial, we need to know the amount of reagent in each vial. The standard AccQ•Tag reagent vial contains 3–4 mg of reagent, which is approximately 10–14 µmoles of reagent. Because the reagent is dissolved in 1 mL of acetonitrile, and 20 µL is used for each 100 µL derivatization reaction, each reaction vessel contains 210–280 nmoles of derivatization reagent.

Because each reaction vial contains 210–280 nmol of reagent, and we need 4–6x molar excess for each sample, there should be no less than 40–140 nmoles of total amines in each reaction.

Example calculation:
For a protein sample, use the weight of sample and the average weight of an amino acid to estimate the required excess.

Step 1: Convert mg/mL concentration of stock to molar concentration of amino acids.

Step 1: Convert mg/mL concentration of stock to molar concentration of amino acids.

For example, with a 1-mg/ml protein concentration, and an average molecular weight of 110, the amount of protein in the sample is determined as follows:

where MW is converted to units from g/mol to µg/µmol, and

    1 mL = 103 µL

    1 mmol = 103 µmol = 106 nmol

Step 2: Determine the molar amount of amino acids in each reaction.

Step 2: Determine the molar amount of amino acids in each reaction.

Once the molar concentration is determined, the amount of amino acids in the sample is calculated.

Using the 1-mg/mL protein in Step 1, which required a 27x dilution (5 µL hydrolysate stock + 135 µL buffer) from Section 6.3.1, the nmol in 10 µL of the diluted sample is calculated as follows:

3.3 nmol is well below the 140 nmol limit, so the sample is acceptable.

6.4 HPLC-BASED AMINO ACID ANALYSIS

6.4 HPLC-BASED AMINO ACID ANALYSIS

The HPLC AccQ•Tag Method, commercialized by Waters Corporation in 1992, uses the same pre-column derivatization step as the AccQ•Tag Ultra Method that was introduced in 2006. The AccQ•Fluor reagent, 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate (AQC), derivatizes primary and secondary amines in a simple, single-step reaction to yield highly stable, fluorescent adducts. We offer the AccQ•Tag Method as a system package that includes prepackaged reagents and extensive documentation. The AccQ•Tag chemistry package contains the items needed for up to 250 analyses of protein and peptide hydrolysate amino acids.

The AccQ•Tag Derivatization Kit contains five sets of the derivatizing reagents. Each set of reagents includes one vial each of the following:

  • AccQ•Fluor Borate Buffer – added to the samples to ensure optimum pH for derivatization.
  • AccQ•Fluor Reagent Powder – 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate. (AQC) derivatizing reagent (shipped dry for maximum stability).
  • AccQ•Fluor Reagent Diluent – acetonitrile, is used to reconstitute the reagent for derivatization.
Figure 11. Representative chromatograms of HPLC-based, amino acid analysis using Waters AccQ•Tag Method.

6.5 UPLC-BASED AMINO ACID ANALYSIS SOLUTION

6.5 UPLC-BASED AMINO ACID ANALYSIS SOLUTION

The Waters UPLC Amino Acid Analysis Solution is holistically designed for turnkey amino acid analyses. Pre-column derivatized amino acids are resolved on a Waters ACQUITY UPLC System using the included AccQ•Tag Ultra, reversed-phase UPLC column, eluents, and methods. Rugged derivatization chemistry, stable chromatographic baselines, and superior amino acid resolution help ensure accurate, precise, and consistent quantitative results.

The UPLC Amino Acid Analysis Solution includes:

  • Waters AccQ•Tag Ultra chemistry supplies, including column, reagents, and eluents – all QC-tested with the amino acid analysis application
  • Empower 2 Software, preconfigured projects, methods, and report formats
  • Full system and application-level support documentation

Waters ACQUITY UPLC System supports three different optical detectors: tunable UV, PDA, and fluorescence detector.

Figure 12. Waters UPLC Amino Acid Analysis Solution.

6.5.1 Accurate amino acid analysis from varied sample matrices

6.5.1 Accurate amino acid analysis from varied sample matrices

The UPLC Amino Acid Analysis Solution includes two complete methods that use the same instrumentation and chemistries. The first is suitable for the amino acids derived from protein hydrolysates. The second is suitable for the larger number of free amino acids found in process samples such as cell culture or fermentation broths. The methods differ in the dilution of the AccQ•Tag Ultra Eluent A and the separation column temperature. There are no user adjustments of pH or modifications of composition for either Eluent A or Eluent B.

Figure 13. Representative chromatograms of UPLC AccQ•Tag Solutions. (A) Separation of standard amino acids using the UPLC Amino Acid Analysis Solution Hydrolysate Method. (B) Separation of the larger set of standard amino acids using the UPLC Amino Acid Analysis Solution Cell Culture Method. No modifications of the mobile phase or composition is required.

6.5.2 Amino acid analysis of hydrolyzed proteins

6.5.2 Amino acid analysis of hydrolyzed proteins

Amino acid analysis of proteins is used both as part of a structural determination and as a measure of the total amount of protein in a sample. The sample is hydrolyzed prior to analysis. For structural analyses, the observed molar ratios of the amino acids are compared to the values expected from the sequence.

For protein amounts, the weights of the amino acids are summed. This measure of protein concentration is used to calculate extinction coefficients where the sample composition interferes with common protein assays. Both weight percent of each amino acid and total protein mass are used to assess the nutritional content of foods and feeds. The Waters UPLC Amino Acid Analysis Solution provides robust, routine analyses across all these applications.

Figure 14. Amino acid analysis of a pure protein hydrolysis.
Figure 15. Amino acid analysis of hydrolyzed poultry diet.

Related

Enable your laboratory to obtain accurate amino acid composition from protein/peptide hydrolysates, physiologic fluids, feeds, foods, pharmaceutical preparations, and myriad other samples using traditional HPLC or UHPLC instrumentation.

Accurately separate, identify, & quantify amino acids in cell culture, protein hydrolysates, foods, and feeds with Waters AccQ•Tag and AccQ•Tag Ultra Amino Acid Analysis standards and kits for HPLC, UHPLC, and UPLC.

Waters offers three distinctively reliable methods for accurate amino acid analyses. All three utilize pre-column derivatization methods followed by reversed-phase chromatography with on-line detection of well resolved peaks using either UV absorbance or fluorescence detection.

Realize the full potential of your automation workstations with Waters sample preparation kits, and seamlessly automate your Waters workflow to gain laboratory efficiency.
Back To Top Back To Top