For forensic toxicology use only.
This application note describes a complete method for the solid phase extraction and UPLC-MS/MS analysis of pain management drugs and drugs of abuse for frensic toxicology.
A number of advantages are highlighted.
Sample preparation is optimized to efficiently extract all analytes with a simplified procedure that reduces the number of manual steps. The water wettable nature of the sorbent enables in-well sample pretreatment and direct loading without conditioning and equilibration, eliminating sample transfer and potential transcription errors. The efficient and reproducible extraction is evident in the high recoveries, consistent matrix effects, and accurate and precise quantitative data.
The use of the ACQUITY UPLC BEH C18 Column results in rapid analysis of a large panel while maintaining all required baseline separations for accurate quantification.
The Waters Xevo TQ-S micro, with features such as StepWave Technology and XDR Detector ensures extremely rapid and accurate quantification of all compounds over wide dynamic ranges. This enables the simultaneous quantification of 6-MAM at 2 ng/mL and methamphetamine at 2500 ng/mL.
This combination of sample preparation, UPLC separation, and MS/MS detection optimizes the workflow and results in a rapid, accurate, and precise method.
Analyte panels for use in forensic toxicology analysis typically include illicit drugs and common drugs of abuse. Often, multiple methods are used to obtain a comprehensive view of the multiple drug classes. These methods may include immunoassay, GC-MS, LC-MS/MS, or a combination of methods. Waters has developed a method for the quantification of a comprehensive drug panel to achieve the appropriate analytical sensitivity, selectivity, and accuracy for unambiguous identification for forensic toxicology.
This method employs a simple sample extraction procedure using Oasis MCX µElution Plates coupled with a rapid and reproducible chromatographic method using an ACQUITY UPLC BEH C18 Column that achieves baseline separation for all critical pairs of potentially interfering analytes. A Waters Xevo TQ-S micro with Xtended Dynamic Range (XDR) capablilities provided the analytical sensitivity and dynamic range capabilities required for this diverse group of compounds.
All standards were obtained from Cerilliant (Round Rock, TX) and Cayman Chemical (Ann Arbor, MI). A mixed stock solution was prepared in methanol at concentrations of 2, 10, and 25 µg/mL, depending upon the analyte. An internal standard stock solution was prepared in methanol at a concentration of 1 µg/mL. Stable isotope labeled internal standards were used for all compounds except naltrexone, methedrone, dehydronorketamine, m-OH-benzoylecgonine, α-Pyrrolidinovalerophenone (alpha-PVP) metabolite 1, meprobamate, flurazepam, norpropoxyphene, and clonazepam. In those cases, either the internal standard interfered with the quantification of one of the other analytes (naltrexone and clonazepam) or the stable labeled IS was not readily available. Samples were prepared by diluting stock solutions into pooled, blank urine. External quality control material was obtained from UTAK Laboratories (Valencia, CA). All analytes, along with their retention times and calibration ranges are listed in Table 1.
LC system: |
ACQUITY UPLC I-Class (FTN) |
Column: |
ACQUITY UPLC BEH C18, 1.7 μm, 2.1 x 100 mm |
Column temp.: |
40 °C |
Sample temp.: |
10 °C |
Injection volume: |
5 μL |
Flow rate: |
0.6 mL/min. |
Mobile phase A (MPA): |
0.1% Formic acid in MilliQ water |
Mobile phase B (MPB): |
0.1% Formic acid in acetonitrile (ACN) |
Purge solvent: |
50:50 MeOH:H2O |
Wash solvent: |
25:25:25:25 MeOH:H2O:IPA:ACN |
Time (min) |
Flow (mL/min) |
% MPA |
% MPB |
---|---|---|---|
0 |
0.6 |
98 |
2 |
3.33 |
0.6 |
33 |
67 |
3.5 |
0.6 |
10 |
90 |
3.6 |
0.6 |
98 |
2 |
4 |
0.6 |
98 |
2 |
MS system: |
Xevo TQ-S micro |
Ionization mode: |
ESI positive |
Desolvation temp.: |
500 °C |
Desolvation gas flow: |
1000 L/hr |
Cone gas flow: |
150 L/hr |
Acquisition range: |
MRM transitions optimized for individual compounds |
Capillary voltage: |
1.0 kV |
Collision energy: |
Optimized for individual compounds (See Appendix 1) |
Cone voltage: |
Optimized for individual compounds (See Appendix 1) |
Data management: |
MS software: MassLynx Quantification software: TargetLynx XS |
Analyte recoveries and matrix effects were calculated as described previously.1 Internal standard corrected matrix effects were calculated using the response factor of the analyte.
100 µL of urine was added to individual wells of an Oasis MCX µElution Plate, followed by 100 µL of a solution containing hydrolysis buffer, 10 µg/mL of β-glucuronidase enzyme, and 100 ng/mL internal standards and mixed by several aspirations. After incubation, 200 µL of 4% H3PO4 was added and mixed by several aspirations. All samples were drawn directly into the sorbent bed by vacuum and subsequently washed with 200 µL of 80:20 H2O:MeOH. The plate was dried under high vacuum (~15 inch Hg) for one minute to remove as much of the wash solution as possible. Samples were eluted using 2 x 25 µL of 50:50 ACN:MeOH containing 5% strong ammonia solution (Fisher, 28–30%). All samples were diluted with 150 µL of sample diluent (2% ACN:1% formic acid in MilliQ water) prior to LC-MS/MS analysis. A graphical workflow of the extraction procedure is shown in Figure 1.
All test compounds are listed in Table 1, along with their retention times and calibration ranges. Figure 2 shows the chromatography of all compounds included in the panel on the ACQUITY UPLC BEH C18 Column. Meprobamate and norpropoxyphene were included in the panel but were only monitored qualitatively, as they are not fully compatible with the sample preparation procedures. As with any multi-analyte panel, care must be taken to ensure that compounds and internal standards do not interfere with each other. Figures 3A and 3B highlight the chromatography of several groups of analytes with the potential to interfere with each other. In each case, either baseline separation is achieved (see naloxone vs. 6-MAM, Figure 3B) or the MRMs do not interfere with each other (see dehydronorketamine and ethylone, Figure 3A). In some cases, certain internal standards were not used.
For example, clonazepam-d4 was not used as it interfered with the quantification of lorazepam. The high efficiency of the UPLC Column enabled all compounds to elute in just over three minutes, without any compromise in resolution for this large panel, with a total run time of four minutes.
The goal of any extraction technique is to achieve efficient and reproducible recovery for all relevant analytes. As in previous work, the wash protocol was modified from the traditional MCX technique to accommodate the benzodiazepines.2 Figure 4 shows the mean extraction recoveries of the entire panel of compounds from six different lots of urine. With the exception of meprobamate and norpropoxyphene, all compounds but two (MDMA and EDDP) had recoveries greater that 70%. Extraction efficiencies were also consistent. Coefficients of variation (%CV) were less than 10% for all quantitative compounds. Recovery data for individual batches followed the same pattern. These highly efficient recoveries across different matrix lots demonstrate the robustness of the extraction technique and are important for quantification of these compounds in samples from different sources.
Matrix effects were also evaluated using multiple lots of urine. As with recovery, consistent matrix effects are essential for accurate quantification. Figure 5A shows the aggregate matrix effects from six lots of urine. Ion suppression was observed for the majority of analytes, with up to 60% ion suppression was observed for morphine and hydroxymorphone. However, with only two exceptions (m-OH BZE and α-OH midazolam) standard deviations from matrix effects were less than 20% indicating consistent matrix lot to lot performance. Figure 5B shows the matrix effects when corrected using the internal standards. In this case 75/78 of the corrected matrix effects were less than 20%.
Seven point calibration curves were extracted across the concentration ranges shown in Table 1. Calibration ranges were tailored to reflect the expected concentrations of various compounds. Quality control samples were prepared at 4 concentrations spanning the range of the calibrators, with the lowest at 1.5x the lowest calibrator and the highest at 75% of the highest calibrator. For most compounds, these QC levels were 15, 75, 250, and 750 ng/mL. The compounds at the lower concentrations had QC levels at 3, 15, 50, and 150 ng/mL and the analytes at the higher concentration range had QC levels at 37.5, 187.5, 625, and 1875 ng/mL. Quantitative method validation involved extracting full curves and QC samples over five different days. Calibration curves were extracted in duplicate and six replicates of QC samples were prepared each day. Control limits for individual calibrators and QC samples were ±15% of target values, with the exception of the lowest points, which were required to be within 20%. Precision limits for QC samples were 20% for the lowest QC point and 15% for the other points. Meprobamate and norpropoxyphene were assessed qualitatively only and were not subject to these controls. A summary of the five independent extractions and analyses met all of these criteria and can be seen in Appendix 2. The majority of compounds were within 10% of their target values with %CVs under 10%. For within batch results, all compounds met the accuracy criteria, and the only compound that had precision results greater than 15% was the high amphetamine QC at 18%.
All calibration curves conformed to FDA bioanalytical method validation requirements,3 which dictate that all calibrators be within 15% of target values except the lowest point, which must be within 20% of its target value and that 75% of calibrators meet this criteria. All compounds met these criteria and all curves had R2 values of 0.99 or greater.
Limits of quantification were defined as those points in which the signal was 5X greater than that of an extracted matrix blank, signal to noise ratios were >10, and both bias and %CV were both less than 20%. To evaluate this, six replicates of the lowest calibrator were extracted in one of the validation batches. All compounds met these criteria.
On instrument stability was also assessed. A single batch was extracted and analyzed five times over an eight day period. Through four days, all compounds met the quantitative validation criteria described above.
In order to assess accuracy, external quality control samples from UTAK Laboratories were evaluated. These results can be seen in Tables 2A–2D. Analytes assessed using external quality control samples included opioids, benzodiazepines, stimulants, and synthetic cathinones. These results show that 91/98 (93%) of the results were within 20% of the target value. The larger deviations for analytes such as fentanyl, norfentanyl, and buprenorphine could be a result of slight errors in the preparation of the master stock mix, as these compounds were spiked using low volumes (20 µL of stock solution). In addition, 7-aminoclonazepam may have stability issues in the urine matrix which could account for its low bias. All results had %RSD values <10%.
This application note describes a complete method for the solid phase extraction and UPLC-MS/MS analysis of illicit drugs and drugs of abuse for forensic toxicology. A number of advantages are highlighted.
This combination of sample preparation, UPLC separation, and MS/MS detection optimizes the workflow and results in a rapid, accurate, and precise method.
720006187, March 2019