SFC Prep 150 AP System
Designed for automated UV and/or MS-triggered purification
The SFC Prep 150 AP System is the first and only commercially available preparative SFC system offering automated sample handling, column switching, fraction collection and tracking in an easy-to-use, open bed format.
The SFC Prep 150 AP System provides routine compound purification for drug discovery labs using the MassLynx Software with FractionLynx Application Manager, offering fast separations, high resolution, and high throughput. The inherent benefits of SFC, combined with Waters’ world-renowned service and support provides users with a robust, cost-effective solution to any purification lab.
Specifications
Maximum pressure range |
300 bar (4350 psi) |
|||
Cooling |
Circulating coolant |
|||
Total flow rate range (CO2 + Co-solvent) |
20.0 to 150 mL/min |
|||
Recovery |
No Ketoprofen detected, therefore carry over is less than 0.01% |
Ambient operating temperature |
15 to 40 °C |
Ambient operating humidity |
20 to 80% relative humidity at 40 °C |
Pressure rating |
10,000 psi (680 bar) |
Maximum operating pressure |
8700 psi (592 bar) |
Flow range |
20 to 200 g/min |
Wetted materials |
316 stainless-steel, 440C, 17-4PH stainless-steel, sapphire, GFPM, PTFE, AL |
Number of solvents |
Four solvents selected by the Solvent Select Valve (SSV) |
|||||||
Solvent conditioning |
Sparging one channel per solvent |
|||||||
Operating flow rate range |
0.50 to 150.00 mL/min, in 0.01 mL/min increments |
|||||||
Maximum operating pressure |
6000 psi up to 100 mL/min, roll-off to 5000 psi at 150 mL/min |
|||||||
Plunger seal wash |
Integral, active, programmable |
|||||||
Delay volume |
<6.5 mL |
|||||||
Compressibility compensation |
Automatic and continuous |
|||||||
Priming |
Self prime from solvent container on floor with dry lines |
|||||||
Compositional accuracy |
±3% absolute (full scale) from 5% to 95%; and from 1 to 150 mL/min (methanol or water step gradient) at 600 psi |
|||||||
Compositional precision |
Retention time variation <0.15 min SD (uracil/caffeine; 1, 5, 20, 50, 150 mL/min; 70:30 water/methanol dial-a mix) |
Sample capacity |
Injection: Up to 15 microplates (up to 384 well) Up to 72 vials Up to 544 test tubes |
Collection: Up to 480 test tubes Up to 12 deep-well plates Up to 128 28 mm vials |
|
Syringe delivery accuracy |
>99.0% at full stroke with 1000 µL syringe |
Carryover |
<0.05% |
Injection volume |
Preparative: 1 mL standard (fixed loop), (0.024 in valve passages) |
System pressure ratings |
6000 psi High pressure side, including injector 100 psi Collection side and syringe/needle wash |
Wash pump |
>30 mL/min flow through standard tubing and needle; two wash solvents. Cycle controlled by “Wash Factor” setting |
Available loops |
1.0 mL (standard preparative) 2.5 mL, 5.0 mL, 12.5 mL, 30.0 mL |
Supported syringes |
1.0 mL (standard) 2.5 mL, 5.0 mL, 12.5 mL |
Injection cycle time |
<30s based on standard wash cycle |
Control software |
MassLynx v4.2 or higher |
Physical/environmental |
Without hood Width: 959 mm (37.75 in) Depth: 629 mm (24.75 in) Height: 1020 mm (40 in) With hood Width: 959 mm (37.75 in) Depth: 629 mm (24.75 in) Height: 1060 mm (41.75 in) |
Default syringe |
5 mL |
Default loop size |
2 mL, (Optional 5 mL, 10 mL) |
Injection volume range |
0.010 mL (10 µL) to 9.5 mL (9500 µL) |
Sample delivery precision |
Less than 1.0% RSD; 0.75 mL injection volume, 2.0 mL loop |
Injection linearity |
R2 greater than 0.995 using 20–75% of loop volume (2 mL loop) |
Sample carryover |
No Ketoprofen detected, therefore carry over is less than 0.01% |
Primary wetted materials |
316 stainless-steel, PEEK, PTFE, borosilicate glass |
The Prep Column Oven is a thermally controlled module designed for use with preparative SFC columns.
Temperature range |
Ambient +5.0 to 70.0 °C |
Maximum heat rate |
Approx 6.0 °C/min |
Temperature accuracy |
±0.5 °C |
Maximum column specification |
1 to 6, 20 x 250 mm columns |
Wavelength range |
190 to 800 nm |
Light source |
Prealigned, Intelligent Technology Deuterium lamp |
Wavelength accuracy |
±1.0 nm |
Linearity range |
<5% at 2 AU (Propylparaben 257 nm, 10 mm cell) |
Optical resolution |
1.2 nm |
Noise, wet |
≤60 μAU (254 nm, 2 Hz, 1 s TC, 3.6 BW res, 10 mm analytical cell) |
Drift, dry |
≤5000 μAU/h (2 h warm-up, constant temperature and humidity at 230 nm, 3.6 BW res, 2 Hz) |
Data rate |
Up to 80 Hz |
Path length |
3 mm (semi prep) |
Cell volume |
18.3 μL (semi prep cell) |
Pressure limit |
6000 psi (408 bar) |
Wetted materials |
316 stainless-steel, PEEK, fused silica |
Line voltage |
100 to 240 VAC |
Number of solvents |
One |
||
Solvent conditioning |
One integrated, vacuum degassing channel |
||
Primary and accumulator check valves |
Passive |
||
Pump seal wash |
Equipped with a wash system to flush the rear of the high pressure seal and the plunger |
||
Compressibility compensation |
Automated and continuous with no user intervention required |
||
Priming |
Automated priming at flow rates of 4 mL/min |
||
Flow accuracy |
1.0% of set flow at 0.5 to 2.0 mL/min. Backpressure: 1000 psi ±20% with degassed MeOH using either a volumetric or mass based method |
||
Flow precision |
<0.075 %RSD or <0.02 minutes SD, whichever is greater based on six replicate injections |
||
Acquisition modes |
Full scan MS Selected Ion Recording (SIR) |
RADAR |
An information-rich acquisition approach that allows you to collect highly specific quantitative data for target compounds while providing the ability to visualize all other components |
Mass range |
30 to 1250 m/z |
Scan speed |
Automatically-optimized for enhanced data quality for acquisition rates of up to e.g. 10 Hz for m/z 100 to 1000 or 20 Hz for m/z 50 to 500 |
Mass accuracy |
Mass accuracy is better than ±0.2 Da over full mass range |
Mass stability |
Mass drift is less than 0.1 Da over a 24 hour period |
Linearity of response |
The linearity of response relative to sample concentration, for a specified compound, is up to four orders of magnitude from the limit of detection |
Ion polarity switching time |
25 ms to switch between positive and negative ion modes |
SIR acquisition rate |
Automatically-optimized for enhanced data quality for acquisition rates of up to 100 Hz |
Number of SIR channels |
Up to 1024 SIR channels (32 functions, 32 channels per function) can be monitored in a single acquisition |
Mass resolution |
Automated mass resolution control (0.7 Da) for constant data quality |
SIR sensitivity (ESI+) |
A 100 pg on column injection (5 μL of 20 pg/μL) of sulfadimethoxine, with a mobile phase flow rate of 800 μL/min will give a chromatographic signal-to-noise for m/z 311 greater than 2000:1 (400:1 with integrated diaphragm backing pump) |
SIR sensitivity (ESI-) |
A 50 pg on column injection (5 μL of 10 pg/μL) of chloramphenicol, with a mobile phase flow rate of 800 μL/min will give a chromatographic signal-to-noise for m/z 321 greater than 300:1 (60:1 with integrated diaphragm backing pump) |
SIR signal-to-noise definition |
Signal is defined as the height of the chromatographic peak of interest and noise is defined as the RMS of a continuous section of the mass chromatogram |
The ABPR valve assembly is motor driven and temperature controlled for cooling during depressurization. A built-in pressure sensor provides closed loop feedback for control and pressure alarm monitoring.
Maximum flow rate |
150 mL/min |
Maximum pressure |
4400 psi (300 bar) |
After exiting the Automated Back Pressure Regulator, this heat exchanger is used to ensure a phase separation in the mobile phase between the CO2 and the co-solvent. This “trailing” Heat Exchanger is used to vaporize the CO2 from the mobile phase so that the Gas Liquid Separator is collecting only a sample-enriched solvent mixture.
Pressure rating |
10,000 psi (680 bar) |
Wetted material |
316 stainless-steel |
Collecting in the same tubes or racks is achieved by eliminating the majority of the carbon dioxide, directing only the liquid phase along with the collected fraction to the collector.
Total flow rate range |
20 to 150 mL/min |
|
Co-solvent max flow range |
75 mL/min |
|
Operational pressure range (Optimal) |
0 to 100 psi (0 to 6.8 bar) 50 psi (3.4 bar) |
|
Wetted materials |
316 stainless-steel, PEEK, borosilicate glass |
|
External control |
MassLynx v4.2 |
|
External communication |
Ethernet and RS-232 communications |
|
Event input/outputs |
Rear panel contact closure and/or TTL inputs/outputs |
|
Operating system |
Windows 10 |
Overview
- Offers automated sample handling, column switching, fraction collection, and tracking in an easy-to-use, open bed format
- Achieve fast separations, high resolution, and high throughput
- Attain routine compound purification for drug discovery labs, using the Waters MassLynx/FractionLynx Application Manager
- Benefit from a robust, cost-effective solution to any purification lab, when combined with world-renowned service and support from Waters
Recommended Use: For drug discovery labs needing fast separations, high resolution, and high throughput in routine compound purification.
Features Header
Automate the purification process
The SFC Prep 150 AP System automates the purification process by linking analytical pre-screen, purification and fraction re-analysis with AutoPurify in an “open access” setting.
SFC systems utilize liquid carbon dioxide (CO2) as its main mobile phase in combination with one or more organic solvents resulting in faster equilibration, lower pressure drops across the column, solvent reduction, and decreased cost per sample. The process is reproducible and applicable to a wide range of compounds relevant in the pharmaceutical, life sciences, chemical materials, food, and environmental markets.
Moving SFC forward
The SFC Prep 150 AP System comes equipped with modules to help with routine compound purification, including:
- QGM Low Pressure mixing Quaternary Gradient Pump: This pump delivers flowrates up to a maximum of 150 mL/min as serves as the co-solvent delivery device for this system.
- P200X CO2 Pump: This is a high-pressure delivery pump having dual stainless-steel heads with a cam driven sapphire piston assembly; self-priming check valves; presser sensor; pressure gauge; brushless motor; and a rupture disc assembly. The pumps design lends itself to control based upon feedback from the pressure sensor and mass flowmeter.
- 3767 Sample Manager: This module provides injection and collection capability in an open bed format. The system includes the fume hood assembly for the 3767 for CO2 management. Injection and collection racks must be ordered for the system. The specific racks required will be determined by your needs and requirements.
- Analytical-2-Prep Column Oven: A thermal control module, the Analytical-2-Prep Column Oven's unique drawer design gives the user unprecedented flexibility, enabling the use of 10 mm, and 20 mm diameter columns simultaneously.
Optional modifier stream injection
The Modifier Stream Injector’s function is to inject one sample at a time directly into the co-solvent flow. This device allows for settable air gaps on either side of the sample, which function as a buffer between the solvent and the sample decreasing dilution. There are two benefits to the MSI:
- The sample is injected into the organic portion of the mobile phase (co-solvent) prior to mixing with CO2. The idea is to mitigate the diluent affect by introducing the sample without affecting the overall strength of the mobile phase, and maintaining the programmed solvent percentages throughout the injection and run. Modifier-stream injection provides improved peak shape and resolution, allowing for larger injection volumes, and higher loading (image 1).
- It allows for Stacked Injections. Stacked injections reduce the time between injection cycles and minimize solvent usage. Stacked injections also significantly improve throughput by utilizing all of the available chromatographic space for continuous separation and purification. Usually, injections are made while an already injected sample is on (or eluting from) the column (image 2).